Philips IntelliBridge System configured with a regional setting which uses a comma for a decimal separator: 866417 IntelliBridge System C.01 (or later) 866418 IntelliBridge System C.01 upgrade (or later) Product Usage: The IntelliBridge System is indicated for use in data collection and clinical information management either directly or through networks with independent bedside devices. The system is not intended for monitoring purposes, nor is the system intended to control any of the clinical devices (independent bedside devices / information systems) it is connected to. The IntelliBridge solution is intended for use in professional healthcare facilities by trained healthcare professionals.
FDA Enforcement
Class III
·Terminated·Philips North America, LLC·June 19, 2019
Philips Patient Information Center iX configured with a regional setting which uses a comma for a decimal separator: 866389 Patient Information Center iX B.01 (or later); 866390 Patient Information Center iX B.01 upgrade (or later) Product Usage: The intended use of the Information Center Software is to display physiologic waves, parameters, and trends, format data for strip chart recordings and printed reports, and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. The Information Center Software provides for the retrospective review of alarms, physiologic waves and parameters from its database. An additional intended use of the Information Center Software is to provide primary annunciation of alarms and configuration and control access for networked telemetry monitors
FDA Enforcement
Class III
·Terminated·Philips North America, LLC·June 19, 2019
Philips Patient Information Center iX configured with a regional setting which uses a comma for a decimal separator: 866389 Patient Information Center iX B.01 (or later); 866390 Patient Information Center iX B.01 upgrade (or later) Product Usage: The intended use of the Information Center Software is to display physiologic waves, parameters, and trends, format data for strip chart recordings and printed reports, and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. The Information Center Software provides for the retrospective review of alarms, physiologic waves and parameters from its database. An additional intended use of the Information Center Software is to provide primary annunciation of alarms and configuration and control access for networked telemetry monitors
FDA Recall
Terminated
·Philips North America, LLC·Product code MHX·April 29, 2019
Philips IntelliBridge System configured with a regional setting which uses a comma for a decimal separator: 866417 IntelliBridge System C.01 (or later) 866418 IntelliBridge System C.01 upgrade (or later) Product Usage: The IntelliBridge System is indicated for use in data collection and clinical information management either directly or through networks with independent bedside devices. The system is not intended for monitoring purposes, nor is the system intended to control any of the clinical devices (independent bedside devices / information systems) it is connected to. The IntelliBridge solution is intended for use in professional healthcare facilities by trained healthcare professionals.
FDA Recall
Terminated
·Philips North America, LLC·Product code N/A·April 29, 2019
GE Healthcare, DASH 3000, 4000, 5000 Patient Monitors The Dash 3000/4000/5000 patient monitor is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use of the system is to monitor physiologic parameter data on adult, pediatric and neonatal patients. The Dash 3000/4000/5000 patient monitor is designed as a bedside, portable, and intra-hospital transport monitor that can operate in all professional medical facilities including but not limited to: emergency department, operating room, post anesthesia recovery, critical care, surgical intensive care, respiratory intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care areas located in hospitals, outpatient clinics, free-standing surgical centers, and other alternate care facilities. Physiologic data includes but is not restricted to: electrocardiogram, invasive blood pressure, noninvasive blood pressure, heart rate, temperature, cardiac output, respiration, pulse oximetry, carbon dioxide, bi-spectral index, impedance cardiography, oxygen, and anesthetic agents as summarized in the operator's manual. The Dash 3000/4000/5000 patient monitor is also intended to provide physiologic data over the UNITY NETWORK TM indirectly to clinical information systems (via our Enterprise Gateway) and allow the user to access hospital data at the point-of-care. The information can be displayed, trended, stored, and printed. The Dash 3000/4000/5000 patient monitor was developed to interface with nonproprietary third party peripheral devices that support serial data outputs.
FDA Recall
Terminated
·GE Healthcare, LLC·Product code MHX·July 5, 2011
Evita V500 Ventilator with Software 2.51.01 and Lower - intended for the ventilation of adult, pediatric and neonatal patients Catalog Number: 8416400
FDA Recall
Terminated
·Draeger Medical, Inc.·Product code CBK·February 14, 2022
Babylog VN500 Ventilator with Software 2.51.01 and Lower- intended for the ventilation neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight Catalog Number: 8417400
FDA Recall
Terminated
·Draeger Medical, Inc.·Product code CBK·February 14, 2022
ONLINE TDM N-acetylprocainamide, Part number 04642520190 N- acetylprocainamide reagent Roche Diagnostics Corporation Indianapolis, IN.
FDA Recall
Terminated
·Roche Diagnostics Operations, Inc.·Product code LAN·February 5, 2010
GE Healthcare, S/5 iCentral and iCentral Client, Related to software licenses L-NET05 and L-NETC05, Datex-Ohmeda Inc., PO Box 7550 Madison, WI 53707-7550, USA, GE Healthcare Finland Oy, Helsinki, Finland PO Box 900. FI-00031 GE. The iCentral maintains the network connections between the Datex-Ohmeda bedside monitors and other network devices in Monitor Network. Network connection consists of hardwire network and/or wireless LAN (WLAN) connection. The iCentral can be used for remote monitor management, storage, printing, viewing, reviewing or otherwise processing of information from several bedside monitors or other network devices. The Dates-Ohmeda Network will be used for patients in the hospital and it is meant for continuous use. The device is to be used by qualified personnel only.
FDA Recall
Terminated
·GE Healthcare·Product code MSX·January 26, 2009
DATEX-OHMEDA NETWORK AND ICENTRAL '05, SALES REVISION 4.3 ( Datek-Ohmeda Network and iCentral (including iCentral Client)) The Datex-Ohmeda Network and iCentral (including iCentral Client) transfers information between networked Datex-Ohmeda devices in the Monitor Network as well as between Monitor Network and networked devices in GE Unity NetworkTM. It also allows information transfer between several iCentrals. Within one Monitor Network it allows a networked device to display, store, print and otherwise process information received from other networked devices. The iCentral maintains the network connections between the Datex-Ohmeda bedside monitors and other networked devices in Monitor Network. Network connections consist of hardwired network cables and/or Wireless LAN (WLAN) connections. Furthermore, it coordinates the transfer of information between devices in the Datex-Ohmeda Network, between the Datex-Ohmeda Network and devices in GE Unity Network as well as between the Datex-Ohmeda Network and Hospital Information Systems (HIS). The iCentral can be used for remote monitor management, storing, printing, viewing, reviewing or otherwise processing of information from several bedside monitors or other networked devices. The iCentral Client can be used for remote monitor management, printing, viewing, reviewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda Network will be used for patients in the hospital and it is meant for continuous use. The device is for use by qualified personnel only.
FDA Recall
Terminated
·Ge Healthcare, Llc·Product code MSX·June 25, 2010
Babi.Plus 12.5 cm H2O Pressure Relief Manifold, REF 1030, RX Only, UDI: 04710810093562
FDA Recall
Terminated
·Galemed Corporation 87 Wu-Jia Li-Gong 2nd Road I-Lan Taiwan·Product code BZD·May 30, 2019
Babi.Plus 12.5 cm H2O Pressure Relief Manifold with Gas Supply Tube and Oxygen Sensor Port, REF 2691, Rx Only, UDI: 04710810100802
FDA Recall
Terminated
·Galemed Corporation 87 Wu-Jia Li-Gong 2nd Road I-Lan Taiwan·Product code BZD·May 30, 2019
Babi.Plus Bubble PAP Valve with 12.5 cm H2O Pressure Relief Manifold, REF 1002, Rx Only, UDI: 24710810093559
FDA Recall
Terminated
·Galemed Corporation 87 Wu-Jia Li-Gong 2nd Road I-Lan Taiwan·Product code BZD·May 30, 2019
ADP2.1 Adapter, non-sterile reusable Model Number: 30-802-1 ,an accessory used with the TissueLink BiPolar Floating Device
FDA Recall
Terminated
·Product code GEI·July 22, 2004
ADP 3.2 Adapter, non sterile reusable Model Number: 30-801-1, an accessory used with the TissueLink Bipolar Floating Device
FDA Recall
Terminated
·Product code GEI·July 22, 2004
Elan Diagnostics CHO ATAC Cholesterol Reagent Kit Product Number: 516-250
FDA Recall
Terminated
·Elan Diagnostics, Inc.·Product code CHH·November 7, 2002
XiO RTP System. Used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed.
FDA Recall
Terminated
·Elekta, Inc.·Product code MUJ·November 25, 2013
Presource PBDS, Kit, No Spiro Diabetes. Kit includes a exerciser volumetric 5000 ML device, two items of cloth, skin prep 2% CHG, a packet of 4 chlorhexidine gluconate antiseptic cloths, and Gatorade pouches. Product Usage: The kit is provided to a patient pre-surgery, as part of the Enhanced Recovery After Surgery protocol. Items included at 1) an exerciser volumetric 5000ML device that is utilized to foster improved lung performance in an effort to avoid potential pulmonary post-operative complications, 2) up to two items of cloth, skin prep 2% CHG, a packet of for chlorhexidine gluconate antiseptic cloths utilized by the patient to reduce the risk of surgical site infections by wiping the body in a prescribed manner, 3) up to two Gatorade to help restore the body s electrolytes when fasting prior to surgery.
FDA Enforcement
Class II
·Terminated·Cardinal Health 200, LLC·July 25, 2018
Presource PBDS, Kit, No Spiro. Kit includes a exerciser volumetric 5000 ML device, two items of cloth, skin prep 2% CHG, a packet of 4 chlorhexidine gluconate antiseptic cloths, and Gatorade pouches. Product Usage: The kit is provided to a patient pre-surgery, as part of the Enhanced Recovery After Surgery protocol. Items included at 1) an exerciser volumetric 5000ML device that is utilized to foster improved lung performance in an effort to avoid potential pulmonary post-operative complications, 2) up to two items of cloth, skin prep 2% CHG, a packet of for chlorhexidine gluconate antiseptic cloths utilized by the patient to reduce the risk of surgical site infections by wiping the body in a prescribed manner, 3) up to two Gatorade to help restore the body s electrolytes when fasting prior to surgery.
FDA Enforcement
Class II
·Terminated·Cardinal Health 200, LLC·July 25, 2018
Presource PBDS, Kit, Common. Kit includes a exerciser volumetric 5000 ML device, two items of cloth, skin prep 2% CHG, a packet of 4 chlorhexidine gluconate antiseptic cloths, and Gatorade pouches. Product Usage: The kit is provided to a patient pre-surgery, as part of the Enhanced Recovery After Surgery protocol. Items included at 1) an exerciser volumetric 5000ML device that is utilized to foster improved lung performance in an effort to avoid potential pulmonary post-operative complications, 2) up to two items of cloth, skin prep 2% CHG, a packet of for chlorhexidine gluconate antiseptic cloths utilized by the patient to reduce the risk of surgical site infections by wiping the body in a prescribed manner, 3) up to two Gatorade to help restore the body s electrolytes when fasting prior to surgery.
FDA Enforcement
Class II
·Terminated·Cardinal Health 200, LLC·July 25, 2018