Babylog VN500 Ventilator with Software 2.51.01 and Lower- intended for the ventilation neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight Catalog Number: 8417400
Recall
- Recall Number
- Z-0871-2022
- Event Number
- 89702
- Firm
- Draeger Medical, Inc.
- FEI Number
- 2517967
- Product Code
- CBK
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- February 14, 2022
- Terminated
- January 7, 2025
- Address
- 3135 Quarry Rd, Telford, PA, 18969-1042
Description
Babylog VN500 Ventilator with Software 2.51.01 and Lower- intended for the ventilation neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight Catalog Number: 8417400
Cybersecurity Improvement Action-not equipped against potential cyber security threats
Draeger issued "Urgent Medical Device Correction Notices" on 2/14/22 to the Director, Clinical Engineering and Risk Manager via FedEx with tracking/confirmation of delivery. Letter states reason for recall, health risk and action to take: Continue to follow the recommendations in the instructions for use: o Limit or control physical access to the abovementioned ventilators. o Do not connect any non-approved devices to the USB, LAN and DVI interfaces. o Be attentive to notifications, alarms, and alerts Consider closing/covering all unused USB, LAN and DVI interfaces Should you decide to close/cover unused interfaces, Draeger offers to provide tools upon request to cover or close these data interfaces of the ventilators free of charge. For the intended and authorized use of the interfaces the USB locks and interface covers can be removed with the provided tool. To order this material please contact Dragger Customer Success via phone between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437 (press 1 at the prompt, the 2), via fax at 1-215-723-5935, or via email at [email protected]. You will need to know the material number to order, which is dependent on the model of the C500 Cockpit that is installed on your V500/VN500 Ventilator. If you have any questions regarding the operation of your Evita V500 or Babylog VN500, please contact Drager Service Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437 (press 2 at the prompt, then 2, then 1).
US Nationwide
13,687 (2,987 units US)