XiO RTP System. Used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed.
Recall
- Recall Number
- Z-0928-2014
- Event Number
- 67224
- Firm
- Elekta, Inc.
- FEI Number
- 1037831
- Product Code
- MUJ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- November 25, 2013
- Posted
- February 4, 2014
- Terminated
- December 2, 2015
- Address
- 400 Perimeter Center Ter Ne, Ste 50, Atlanta, GA, 30346-1227
Description
XiO RTP System. Used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed.
When bolus is present, the Effective Depth (with bolus) to the weight point returned on the Source Data Report is incorrect.
Corrective Action 1 - An important safety notice, LINXIO0018 - Incorrect Effective Depth for Bolused Beams is in the process of being distributed to affected customers. (Release date November 25, 2013). Corrective Action 2 - Develop a fix to the defect in an expeditious timeframe and make available in Patch Release 4.0.03 and Release 5.00.01.
Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AL, AK, AZ, CA, CO, CT, DE, FL, GA, IL, IN, KS, KY, LA, ME, MD, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, QA, WA, WV, WI, and WY, and the countries of Albania, Algeria, Argentina, Austria, Australia, Bosnia, Bulgaria, Belgium, Bahrain, Brazil, Belarus, Botswana, Bangladesh, Canada, China, Columbia, Costa Rica, Croatia, Cuba, Cyprus, Chile, Czech Republic, Ecuador, Egypt, Estonia, France, Finland, Gabon, Germany, Great Britain, Greece, Guatemala, Hungary, India, Indonesia, Italy, Ireland, Israel, Iraq, Jordan, Japan, Kosovo, Kuwait, Lithuania, Libya, Latvia, Morocco, Malaysia, Malta, Mexico, Myanmar, Netherland, New Zealand, Pakistan, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, Slovakia, South Africa, South Korea, Spain, Sudan, Syria, Taiwan, Thailand, Turkey, Turkmenistan, Ukraine, Venezuela, and Vietnam.