FDA Enforcement Class III Terminated

Philips IntelliBridge System configured with a regional setting which uses a comma for a decimal separator: 866417 IntelliBridge System C.01 (or later) 866418 IntelliBridge System C.01 upgrade (or later) Product Usage: The IntelliBridge System is indicated for use in data collection and clinical information management either directly or through networks with independent bedside devices. The system is not intended for monitoring purposes, nor is the system intended to control any of the clinical devices (independent bedside devices / information systems) it is connected to. The IntelliBridge solution is intended for use in professional healthcare facilities by trained healthcare professionals.

Recall: Z-1772-2019 · Reported June 19, 2019

Enforcement

Recall Number
Z-1772-2019
Event ID
82914
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Philips North America, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 19, 2019
Initiation Date
April 29, 2019
Classification Date
June 12, 2019
Termination Date
November 22, 2024
Address
3000 Minuteman Rd, N/A, Andover, MA, 01810-1032, United States

Description

Philips IntelliBridge System configured with a regional setting which uses a comma for a decimal separator: 866417 IntelliBridge System C.01 (or later) 866418 IntelliBridge System C.01 upgrade (or later) Product Usage: The IntelliBridge System is indicated for use in data collection and clinical information management either directly or through networks with independent bedside devices. The system is not intended for monitoring purposes, nor is the system intended to control any of the clinical devices (independent bedside devices / information systems) it is connected to. The IntelliBridge solution is intended for use in professional healthcare facilities by trained healthcare professionals.

Reason

Infusion Pump Data Storage Accuracy-Data from the BBraun Space LAN or Arcomed UniqueDoc infusion pumps transmitted via the HL7 output interface through the Patient Information Center iX using a LAN driver may be recorded in the patients chart or electronic medical record at exactly 100 times the actual bolus rate, infusion rate and total volume values

Code Info

866417 IntelliBridge System C.01 (or later) 866418 IntelliBridge System C.01 upgrade (or later

Distribution

Worldwide Distribution - US Nationwide Foreign: Andorra Aruba Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Costa Rica Croatia Czech Republic Denmark Dominican Rep Egypt Estonia Finland France French Guiana Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iran Ireland Israel Italy Japan Jordan Kenya Korea Rep Lao Latvia Lebanon Lithuania Luxembourg Macedonia Malaysia Maldives Martinique Mauritius Mayotte Mexico Mongolia Morocco Myanmar Namibia Nepal Netherlands New Caledonia New Zealand Norway Oman Pakistan Palestine Panama Peru Philippines Poland Portugal Puerto Rico Qatar R¿union Romania Russia Russian Fed. Saudi Arabia Serbia Singapore Slovenia South Africa South Korea Spain Sri Lanka Sweden Switzerland Taiwan Thailand Turkey United Kingdom Uruguay Utd Arab Emir Utd.Arab.Emir. Uzbekistan Viet Nam Vietnam White Russia

Quantity

180 units