The Synchro Neuro Guidewires are a steerable guidewire family with a shapeable tip. The outside diameter of the Synchro 10 guidewire is 0.010in. The outside diameter of the Syncrho 14 guidewire is 0.014in. The guidewires are compatible with existing microcatheters used in common procedures such as those used in endovascular diagnosis and therapy of neurovascular disease. The distal portion of the guidewire tip is radiopaque. For lubricity, the distal portion of the device is coated with a hydrophilic polymer and the proximal portion of the guidewire is coated with polytetrafluoroethylene (PTFE). The torque device included with the guidewire attaches to the proximal end of the wire and functions as a steering guide. Rotation of this device facilitates guidewire placement into the appropriate vessel by precise directional manipulation of the guidewire tip. The introducer included with the guidewire is intended to aid insertion of the guidewire into the catheter hub and/or hemostasis valve.

FDA Enforcement
Class II ·Ongoing·Stryker Neurovascular·June 12, 2024

Discovery NM 630, dual detector free-geometry integrated nuclear imaging system. Models H3100RC, H3100RD, H2401LW. Product Usage: The GE Discovery NM 630 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and in assessment of organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM 630 system is a Nuclear Medicine (NM) system, which is intended to yield General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The scanning modes include planar mode (Static, Multi-gated, Dynamic and Whole body scanning) and tomographic mode (SPECT, Gated SPECT, Whole body SPECT). The acquisition types include single and multi-isotope/ multi peak frame/list mode single-photon imaging. The imaging-enhancement features include assortment of collimators, gating by physiological signals, and real-time automatic body contouring. The GE Discovery NM 630 system may include signal analysis and display equipment, patient and equipment supports, components and accessories. The system may be used for patients of all ages.

FDA Enforcement
Class II ·Terminated·GE Healthcare, LLC·November 19, 2014

Software-Aided Adjunctive Diagnostic Device For Use By Healthcare Providers In Skin Wound Assessment.

FDA classification
FDA Class 2 ·Software-Aided Adjunctive Diagnostic Device For Use By Healthcare Providers In Skin Wound Assessment.

Natus neurology Dantec KEYPOINT.NET 2.20. Catalog/Part Numbers: 9031A070103, 9031A006502, 9033A071103, 9033A072102, 9031A070102, 9033A007402, 9033A071102, 9033A072103, 9033A0701, 9033A0711HA, 9033A071101, 9031A070101, 9031A006401, 9031A006501, 9033A004701, 9031A0032, 9033A0517, 9033A072101, 9031A0031, 9031A0701BV, 9033A0044, 9033A0073AA, 9031A0063, 9033A0523, 9033A071143, 9033A072143, 9031A006406, 9031A006506, 9031A070106, 9033A004606 9033A004706, 9033A007406, 9033A0537, 9033A0544, 9033A071106, 9033A072106, 9033A0524, 9033A0711BV, 9033A007401, 9033A0516, 9031A006403, 9031A006503, 9033A056403, 9033A0721BV, 9031A006402, 9031A006405, 9031A051306, 9031A070105, 9033A0072, 9033A007301, 9033A0711AA, 9033A0721AA Product Usage: Dantec KEYPOINT.NET Software 2.20 is installed on the Keypoint EMG System which is intended as an electro-physiological aid to assess diagnosis and prognosis and to monitor diseases of the central and peripheral nervous system. It can also be used to study functional aspects of nerves and muscles.

FDA Enforcement
Class II ·Terminated·Alpine Biomed Aps·June 5, 2013

Software-Aided Adjunctive Diagnostic Device For Use By Physicians On Lesions Suspicious For Skin Cancer

FDA classification
FDA Class 2 ·Software-Aided Adjunctive Diagnostic Device For Use By Physicians On Lesions Suspicious For Skin Cancer

GE Healthcare Discovery NMCT 670, Discovery NM 630, Optima NMCT 640, & Brivo NM 615. The GE Discovery NMJCT 670 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce: NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The scanning modes include planar mode (Static, Multi-gated, Dynamic and Whole body scanning) and tomographic mode (SPECT, Gated SPECT, Whole body SPECT). The acquisition types include single and multi-isotope/multi peak frame/list mode single-photon imaging. The imaging-enhancement features include assortment of collimators, gating by physiological signals, and real-time automatic body contouring. CT System: Cross sectional images of the body by computer reconstruction of X-Ray transmission data taken at different angles and planes, including Axial, Cine, Helical, Cardiac, and Gated acquisitions. These images may be obtained with or without contrast. The CT system is indicated for head, whole body, cardiac and vascular X-Ray Computed Tomography applications NM + CT System: Combined, hybrid SPECT and CT protocols, for CT-based SPECT attenuation corrected imaging as well as functional and anatomical mapping imaging (localization, registration and fusion). The GE Discovery NM/CT 670 system may include signal analysis and display equipment, patient and equipment supports, components and accessories. The system may include data and image processing to produce images in a variety of trans-axial and reformatted planes. The images can also be post processed to produce additional images, imaging planes, and analysis results. The system may be used for patients of all ages. The GE Discovery NM 630 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and in assessment of organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM 630 system is a Nuclear Medicine (NM) system, which is intended to yield General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The scanning modes include planar mode (Static, Multi-gated, Dynamic and Whole body scanning) and tomographic mode (SPECT, Gated SPECT, Whole body SPECT). The acquisition types include single and multi isotope/multi peak frame/list mode single-photon imaging. The imaging-enhancement features include assortment of collimators, gating by physiological signals, and real-time automatic body contouring. The GE Discovery NM 630 system may include signal analysis and display equipment, patient and equipment supports, components and accessories. The system may be used for patients of all ages. - The GE Optima NM/CT 640 system is intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine (NM) system using a variety of scann

FDA Enforcement
Class II ·Terminated·GE Healthcare, LLC·November 27, 2013

Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX Kit/RE4001-01, Leader TRUE METRIX Kit/RE4002-01, Leader TRUE METRIX Meter Only/RE4002-40, CVS TRUE METRIX Kit/RE4007-01SB, GNP TRUE METRIX Meter Only/RE4011-00, GNP TRUE METRIX Kit /RE4011-01, MEIJER TRUE METRIX Kit/RE4019-01, FRED'S TRUE METRIX Kit/RE4023-01, DISCOUNT DRUG TRUE METRIX Kit/RE4025-01, KINRAY TRUE METRIX Kit/RE4027-01, HEB TRUE METRIX Kit/RE4031-01, McKesson TRUE METRIX Starter Kit/RE4051-43, SUNMARK TRUE METRIX Meter Only/RE4054-00, SUNMARK TRUE METRIX Kit/RE4054-01, Signature Care TRUE METRIX Kit/RE4060-01, CareOne TRUE METRIX Kit/RE4061-01, RITE AID TRUE METRIX Kit/RE4066-01, Liberty TRUE METRIX Meter Only/RE4078-40, HUMANA TRUE METRIX Starter Kit/RE4081-03, Wegmans TRUE METRIX Kit/RE4087-01, HEALTHMART TRUE METRIX Meter Only/RE4089-00, HEALTHMART TRUE METRIX Kit/RE4089-01, ARRIVA TRUE METRIX Meter Only/RE4095-40, HealthyAccents TRUE METRIX Kit/RE4097-01, Shopko TRUE METRIX Kit/RE4098-01, Henry Schein TRUE METRIX Kit/RE4099-40, Henry Schein TRUE METRIX Meter Only/RE4099-45, Millennium Pharmay Services TRUE METRIX Kit/RE4202-43, Kroger TRUE METRIX Kit/RE4203-01, TOPCARE TRUE METRIX Kit/RE4209-01, Foster and Thrive TRUE METRIX Kit/RE4211-01, TRUE METRIX Meter Only/RE4H01-00, TRUE METRIX Kit/RE4H01-01, TRUE METRIX Starter Kit/RE4H01-04, TRUE METRIX NFRS Meter Only/RE4H01-40, TRUE METRIX NFRS Starter Kit/RE4H01-43, TRUE METRIX Kit/RE4H01-51, TRUE METRIX Meter Only (mg/dL)/RE4i01-00, TRUE METRIX Kit (mg/dL)/RE4i01-01, TRUE METRIX Kit (mg/dL)/RE4i01-02, TRUE METRIX Starter Kit (mg/dL)/RE4i03-02, TRUE METRIX Kit (mmol/L)/RE4i03-11, Farmacia Benavides TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i05-00, Farmacia del Ahorro TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i07-00, TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i23-00, TRUE METRIX Meter Only (Jamaica - mmol/L)/RE4i29-11, TRUE METRIX Kit (mg/dL)/RE4i61-02, TRUE METRIX Kit (mg/dL)/RE4i61-02THI, TRUE METRIX Kit (United Kingdom - mmol/L)/RE4i82-11, TRUE METRIX Starter Kit (United Kingdom - mmol/L)/RE4i82-12.

FDA Enforcement
Class I ·Ongoing·Trividia Health, Inc.·March 18, 2026

Vectris(TM) Sure Scan(R) MRI Lead Kit for Spinal Cord Stimulation, Models: (a) 977A160 (b) 977A175 (c) 977A190 (d) 977A260 (e) 977A275 (f) 977A290 Product Usage: The Medtronic Vectris lead kits contain the implantable spinal cord stimulation lead, as well as additional components used for lead implant as part of a Neurostimulation system for pain therapy. A Medtronic implantable Neurostimulation system is indicated for spinal cord stimulation (SCS) systems as an aid in the management of chronic, intractable pain of the trunk and/or limbs-including unilateral or bilateral pain associated with the following conditions: Failed Back Syndrome (FBS) or low back syndrome or failed back; Radicular pain syndrome or radiculopathies resulting in pain secondary to FBS or herniated disk; Post laminectomy pain; Multiple back operations; Unsuccessful disk surgery; Degenerative Disk Disease (DDD)/herniated disk pain refractory to conservative and surgical interventions; Peripheral causalgia; Epidural fibrosis; Arachnoiditis or lumbar adhesive arachnoiditis; Complex Regional Pain Syndrome (CRPS), Reflex Sympathetic Dystrophy (RSD), or causalgia.

FDA Enforcement
Class II ·Terminated·Medtronic Neuromodulation·July 25, 2018

Vectris(TM) Trial Screening Lead Kit for Spinal Cord Stimulation, Models: (a) 977D160 (b) 977D260 Product Usage: The Medtronic Vectris lead kits contain the implantable spinal cord stimulation lead, as well as additional components used for lead implant as part of a Neurostimulation system for pain therapy. A Medtronic implantable Neurostimulation system is indicated for spinal cord stimulation (SCS) systems as an aid in the management of chronic, intractable pain of the trunk and/or limbs-including unilateral or bilateral pain associated with the following conditions: Failed Back Syndrome (FBS) or low back syndrome or failed back; Radicular pain syndrome or radiculopathies resulting in pain secondary to FBS or herniated disk; Post laminectomy pain; Multiple back operations; Unsuccessful disk surgery; Degenerative Disk Disease (DDD)/herniated disk pain refractory to conservative and surgical interventions; Peripheral causalgia; Epidural fibrosis; Arachnoiditis or lumbar adhesive arachnoiditis; Complex Regional Pain Syndrome (CRPS), Reflex Sympathetic Dystrophy (RSD), or causalgia.

FDA Enforcement
Class II ·Terminated·Medtronic Neuromodulation·July 25, 2018

Assay To Measure Pct To Aid In The Risk Assessment Of Critically Ill Patients On Their First Day Of Icu Admission

FDA classification
FDA Class 2 ·Assay To Measure Pct To Aid In The Risk Assessment Of Critically Ill Patients On Their First Day Of Icu Admission

Brand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/A Product Description: Preface: This report under 21 CFR 806.10 is being submitted for the dV5 system tower, as the NIR Handheld Camera or SOCK is developed and integrated as part of the dV5 system. The affected component of the dV5 system tower is the NIR handheld camera, and is supplied by Schoelly as the manufacturer. Therefore, the product description in this report references the NIR handheld camera system. Product Description: The NIR handheld camera system is compatible for use with the da Vinci¿ 5 system to aid in surgery. The components of the NIR handheld camera system include a camera control unit and a light source, NIR handheld camera, and additional components that together, connect to the Tower of the da Vinci 5 system. When setting up the da Vinci 5 system, the image from the NIR handheld camera system is displayed on the Tower monitor. Component: Yes, the SOCK is the component and integrated into the dV5 system

FDA Enforcement
Class II ·Ongoing·Intuitive Surgical, Inc.·June 10, 2026

HeartMate 3 Left Ventricular Assist System (LVAS): 1. HeartMate 3 Sealed Outflow Graft with Bend Relief, Investigational Device Exemption (IDE) 2. HeartMate 3 Sealed Outflow Graft with Bend Relief, US 3. HeartMate 3 Sealed Outflow Graft with Bend Relief, OUS 4. HeartMate 3 LVAS Implant Kit, IDE 5. HeartMate 3 LVAS Implant Kit, US 6. HeartMate 3 LVAS Implant Kit, OUS HeartMate 3 Left Ventricular Assist System: The HeartMate 3 LVAS is designed to provide therapeutic benefit to those affected with advanced heart failure. The LVAS assumes some or all of the workload of the left ventricle. The LVAS features a Left Ventricular Assist Device (LVAD), a blood pump intended for long-term implantation in such patients, an extracorporeal Controller, plus supporting equipment.

FDA Enforcement
Class I ·Ongoing·Thoratec Corp.·March 27, 2024