FDA Enforcement Class II Ongoing

The Synchro Neuro Guidewires are a steerable guidewire family with a shapeable tip. The outside diameter of the Synchro 10 guidewire is 0.010in. The outside diameter of the Syncrho 14 guidewire is 0.014in. The guidewires are compatible with existing microcatheters used in common procedures such as those used in endovascular diagnosis and therapy of neurovascular disease. The distal portion of the guidewire tip is radiopaque. For lubricity, the distal portion of the device is coated with a hydrophilic polymer and the proximal portion of the guidewire is coated with polytetrafluoroethylene (PTFE). The torque device included with the guidewire attaches to the proximal end of the wire and functions as a steering guide. Rotation of this device facilitates guidewire placement into the appropriate vessel by precise directional manipulation of the guidewire tip. The introducer included with the guidewire is intended to aid insertion of the guidewire into the catheter hub and/or hemostasis valve.

Recall: Z-1939-2024 · Reported June 12, 2024

Enforcement

Recall Number
Z-1939-2024
Event ID
94543
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Stryker Neurovascular
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 12, 2024
Initiation Date
April 18, 2024
Classification Date
May 31, 2024
Address
47900 Bayside Pkwy, N/A, Fremont, CA, 94538-6515, United States

Description

The Synchro Neuro Guidewires are a steerable guidewire family with a shapeable tip. The outside diameter of the Synchro 10 guidewire is 0.010in. The outside diameter of the Syncrho 14 guidewire is 0.014in. The guidewires are compatible with existing microcatheters used in common procedures such as those used in endovascular diagnosis and therapy of neurovascular disease. The distal portion of the guidewire tip is radiopaque. For lubricity, the distal portion of the device is coated with a hydrophilic polymer and the proximal portion of the guidewire is coated with polytetrafluoroethylene (PTFE). The torque device included with the guidewire attaches to the proximal end of the wire and functions as a steering guide. Rotation of this device facilitates guidewire placement into the appropriate vessel by precise directional manipulation of the guidewire tip. The introducer included with the guidewire is intended to aid insertion of the guidewire into the catheter hub and/or hemostasis valve.

Reason

Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.

Code Info

Catalog/UDI-DI/Lots: M00316310/07613252187158/0000131357, 0000132153, 0000132740, 0000135026, 0000138147, 0000138148, 0000138579, 0000142533, 0000144961, 0000145874, 0000154419, 0000156324, 0000156614, 0000158907, 0000158911, 0000164603, 0000173346, 0000173701, 0000173702, 0000173771, 0000173995; M00316330/07613252187165/0000142692, 0000158995; M00313010/07613252186922/0000129027, 0000129028, 0000129029, 0000129218, 0000129219, 0000129630, 0000129631, 0000129841, 0000129842, 0000129902, 0000130225, 0000130226, 0000130410, 0000130411, 0000130412, 0000130502, 0000130503, 0000130643, 0000130645, 0000130869, 0000130870, 0000131304, 0000131307, 0000131343, 0000131724, 0000131725, 0000131726, 0000132154, 0000132155, 0000132582, 0000132640, 0000132641, 0000133182, 0000133183, 0000133391, 0000133565, 0000133566, 0000133567, 0000133568, 0000133928, 0000133929, 0000134128, 0000134129, 0000134130, 0000134307, 0000134308, 0000135009, 0000135148, 0000135149, 0000135303, 0000135304, 0000135806, 0000135807, 0000136252, 0000136923, 0000137267, 0000137322, 0000137373, 0000137374, 0000137375, 0000138751, 0000139437, 0000141635, 0000142248, 0000142249, 0000142541, 0000142691, 0000143123, 0000143350, 0000143351, 0000143585, 0000143955, 0000144141, 0000144320, 0000144379, 0000144380, 0000144739, 0000144760, 0000144761, 0000144764, 0000144959, 0000144960, 0000145884, 0000145885, 0000146228, 0000146390, 0000146515, 0000146843, 0000146916, 0000147040, 0000147186, 0000147306, 0000147440, 0000147585, 0000147876, 0000148965, 0000149562, 0000150095, 0000150872, 0000150873, 0000151381, 0000151461, 0000151677, 0000151847, 0000151848, 0000152230, 0000152231, 0000152375, 0000152522, 0000152977, 0000153282, 0000153370, 0000153970, 0000154140, 0000154302, 0000154303, 0000154779, 0000154780, 0000154894, 0000154895, 0000155020, 0000155068, 0000156317, 0000156321, 0000156615, 0000156626, 0000156794, 0000156800, 0000157226, 0000157227, 0000157892, 0000157894, 0000158127, 0000158128, 0000158131, 0000158339, 0000158340, 0000159324, 0000159932, 0000159997, 0000159998, 0000160368, 0000160369, 0000160434, 0000160435, 0000160893, 0000160894, 0000160895, 0000161061, 0000161062, 0000161063, 0000162104, 0000162105, 0000162106, 0000162896, 0000162991, 0000162992, 0000163716, 0000163967, 0000163968, 0000164443, 0000164470, 0000165356, 0000165363, 0000165364, 0000166540, 0000166542, 0000166544, 0000170149, 0000170246, 0000170940, 0000171065, 0000171235, 0000172266, 0000172267, 0000172381, 0000172787, 0000172865, 0000172866, 0000173279, 0000174292, 0000174373, 0000174380, 0000174409, 0000174680, 0000174681, 0000175067, 0000175068, 0000175069, 0000175640, 0000175641, 0000175983, 0000175984, 0000175985, 0000175986, 0000176344, 0000176345, 0000176504, 0000176523, 0000176830, 0000176836, 0000177535, 0000178075, 0000178076, 0000178532, 0000178533, 0000178535, 0000178626, 0000179628, 0000179629, 0000180361, 0000180738, 0000180800, 0000193546; M00313020/07613252186939/0000159372, 0000172593; M00313310/07613252186946/0000135007, 0000135008, 0000136409, 0000136410,0000136679, 0000139937, 0000139938, 0000140307, 0000140689, 0000140690, 0000141634, 0000141833, 0000142246, 0000152799, 0000153160, 0000153453, 0000153968, 0000153969, 0000154301, 0000157222, 0000157223, 0000157224, 0000159320, 0000159370, 0000160740, 0000172380, 0000172594, 0000172619, 0000172661, 0000172662, 0000172748, 0000172786, 0000176476, 0000176840, 0000176841, 0000177540, 0000178232, 0000180657, 0000180737, 0000180795; M00313410/07613252186960/0000173997

Distribution

Worldwide - US Nationwide distribution including in the states of TX, NY, MD, CA, KY, WI, OR, NE, SC, MA, MO, CT, ID, PA, TN, FL, NC, AL, IA, OK, VA, IL, OH, NH, MN, ND, KS, NJ, WV, IN, GA, HI, AR, UT, LA, MI, WA, NV, CO, AZ, DC, VT, ME, DE, AK, WY, MT, NM, SD, MS, RI, GU and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, BULGARIA, CANADA, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, ICELAND, INDIA, IRAN, IRELAND, ISRAEL, ITALIA, JAPAN, JORDAN, KUWAIT, LITHUANIA, LUXEMBOURG, MEXICO, NETHERLANDS, NORTHERN IRELAND, NORWAY, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UAE, UK.

Quantity

105,286