Stryker Orthopaedics Modular Handle Triathlon Instruments, Non-Sterile, REF 6541-4-808, Lot 123, NDC M546654148080S Product Usage: Is a reusable instrument that is part of the Triathlon Primary total knee arthroplasty (TKA) system as well as the Triathlon Partial Knee Replacement (PKR) system. The use of the handles in Primary TKA is optional and may be used to: (1) Assist in stabilizing the Triathlon Femoral Sizer, (2) Aid in stabilization during use and removal of the Triathlon 4:1 Cutting Block and (3) Aid in assembly of the Anterior MIS Distal Resection Guide to the Anterior Skim Cut Guide. Per the Triathlon PKR surgical protocol, the handle is required for a PKR surgery and is used to: (1) Assemble and disassemble the Femoral Trial Drill Guide into the Femoral Trial and (2) Remove the Peg Trial from the Femoral Trial.
Recall
- Recall Number
- Z-2377-2016
- Event Number
- 74634
- Firm
- Stryker Howmedica Osteonics Corp.
- FEI Number
- 3003070421
- Product Code
- JWH
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- June 23, 2016
- Terminated
- March 27, 2019
- Address
- 325 Corporate Dr, Mahwah, NJ, 07430-2006
Description
Stryker Orthopaedics Modular Handle Triathlon Instruments, Non-Sterile, REF 6541-4-808, Lot 123, NDC M546654148080S Product Usage: Is a reusable instrument that is part of the Triathlon Primary total knee arthroplasty (TKA) system as well as the Triathlon Partial Knee Replacement (PKR) system. The use of the handles in Primary TKA is optional and may be used to: (1) Assist in stabilizing the Triathlon Femoral Sizer, (2) Aid in stabilization during use and removal of the Triathlon 4:1 Cutting Block and (3) Aid in assembly of the Anterior MIS Distal Resection Guide to the Anterior Skim Cut Guide. Per the Triathlon PKR surgical protocol, the handle is required for a PKR surgery and is used to: (1) Assemble and disassemble the Femoral Trial Drill Guide into the Femoral Trial and (2) Remove the Peg Trial from the Femoral Trial.
Stryker Orthopaedic is voluntarily recalling the Triathlon Modular Handle because analysis revealed that the press fit specifications between the dowel pin and the mating hole in the shaft were not being achieved potentially resulting in disassociation from the instrument.
Stryker sent an Urgent Product Recall letter dated June 24, 2016 to Branches/Agencies by e-mail and Business Reply Forms was sent to Branches/Agents/Hospital Risk Management via UPS (with return receipt) on June 24, 2016. The letter identified the affected product, problem and actions to be taken. For questions call (201) 831-6693.
Worldwide Distribution - US Nationwide
12,469 units