FDA Recall Open, Classified

Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual

Recall: Z-2340-2025 · Initiated July 8, 2025

Recall

Recall Number
Z-2340-2025
Event Number
96916
Firm
Boston Scientific Neuromodulation Corporation
FEI Number
3006630150
Product Code
NHL
Status
Open, Classified
Root Cause
Component change control
Initiated
July 8, 2025
Posted
August 19, 2025
Address
25155 Rye Canyon Loop, Valencia, CA, 91355-5004

Description

Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual

Reason

Resistance and difficulties could occur while closing the retaining clip locking mechanism, which is part of a burr hole cover kit, which is part of a Deep Brain Stimulation (DBS) system, which may lead to procedural delay, or lead-migration-which may require additional surgery.

Action

On 7/8/2025, correction notices were mailed, emailed, or delivered to healthcare professionals who were asked to do the following: 1. Review the updates to the Surgical Implant Manual stated in this notice. These updates will be found in the IFU once implemented. 2. If you are a facility that has sent products to another hospital or a facility within your network, ensure this notification is forwarded to them. 3. If you are a distributor, this notification must be forwarded to your customers to ensure this notification is carried out to the end-user level. 4. To provide awareness of this information, share this letter with any other clinicians in your hospital who use these devices. 5. If users encounter resistance or difficulties while closing the Retaining Clip locking mechanism, refer to the Surgical Implant Manual updates provided. 6. Maintain a copy of this letter in your facility s records. 7. Complete and return the acknowledgement form via email to [email protected] The Surgical Implant Manual has updates that include: - If you have difficulty locking the Slider, you may need to remove the Lead Stylet before locking the Slider in place, or replace the Retaining Clip using another Burr Hole Cover Kit or Burr Hole Cover Spares Kit. - You may confirm the Slider is fully locked. - Verify that the DBS Lead has not moved from the desired location (e.g. by intraoperative imaging). Report all device-related incidents or quality concerns experienced with the use of these devices to firm at [email protected]

Distribution

Worldwide - US Nationwide distribution including in the states of MN, FL, CO, MD, NY, NE, PA, WI, AZ, TN, MO, TX, MI, MT, MA, WV, AL, VA, NC, CA, OR, WA, LA, OH, NJ, NH, SD, GA, CT, NV, UT, IN, ID, IL, ME, SC, DC, OK, KY, AK, HI, RI, AR, KS, NM and the countries of Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Iran, Ireland, Israel, Italy, Jordan, Kazakhstan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Morocco, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, T¿rkiye, UAE, Ukraine, United Kingdom, Canada, Japan, Argentina, Brazil, Chile, Colombia , Ecuador, Mexico, Hong Kong, China, India, Singapore, Korea, Taiwan.

Quantity

24,583