Product Code: NHL FDA class 3

Stimulator, Electrical, Implanted, For Parkinsonian Symptoms

View full classification →
Adverse events in period
426
-44% vs. prior period (761)
Deaths reported
9
Recalls in period
2
Class I enforcement
0

Adverse events per month

2024-07 – 2026-06
Loading...

Events by type

Period vs. prior period
Loading...
Event type
Period
Prior
Death
9
20
Injury
351
660
Malfunction
65
80
Other
1
1

Most reported coded problems

Top 15
Product problems
Count
Adverse Event Without Identified Device or Use Problem
197
Unexpected Therapeutic Results
106
High impedance
56
Battery Problem
33
Charging Problem
25
Wireless Communication Problem
21
Migration
21
Inappropriate/Inadequate Shock/Stimulation
16
Use of Device Problem
13
Fracture
12
Intermittent Energy Output
11
Off-Label Use
10
Premature Discharge of Battery
7
Nonstandard Device
7
Appropriate Device Problem Term/Code Not Available
7
Patient problems
Count
No Clinical Signs, Symptoms or Conditions
103
Insufficient Information
67
Swelling/ Edema
59
Bacterial Infection
45
Movement Disorder
43
Shaking/Tremors
39
Inflammation
38
Pain
29
Undesired Nerve Stimulation
28
Unspecified Infection
24
Erythema
23
Wound Dehiscence
22
Erosion
22
Discomfort
21
Hemorrhage/Blood Loss/Bleeding
20

Recalls in period

2 total
FDA enforcement classification: Class II: 2
Date
Recalling firm
Status
2025-07-08
Open, Classified
2025-07-08
Open, Classified

Adverse events by year

Loading...

Methodology & data provenance

For your PSUR appendix

Search protocol: FDA MAUDE adverse event reports, FDA device recalls, and FDA enforcement reports (openFDA datasets) were queried for product code NHL, reporting period 2024-07 to 2026-06 (prior comparison period 2022-07 to 2024-06).

Data coverage: adverse event data for these product codes extends through 2026-05. Source openFDA export dated 2026-06-30. Recall records with missing or implausible initiation dates (0 for these codes) are excluded from period counts.

Generated: 2026-07-06 00:35 UTC by BEUDAMED (beudamed.com/fda/psur-report). Dataset sizes: 25,039,198 adverse events, 58,607 recalls, 39,365 enforcement reports.

PSUR database searches, explained

Under the EU MDR, manufacturers must produce a Periodic Safety Update Report (PSUR) (Class IIa and above; a PMS report for Class I) on a defined schedule. A required part of that report is a survey of publicly available vigilance data — most importantly the FDA's MAUDE adverse event database and FDA recall records — covering both your own device and similar devices on the market (the "state of the art").

Doing this by hand means repeated MAUDE queries, manual export, deduplication, and trending — hours of work per report, repeated every reporting period. This tool runs the same search protocol in seconds and documents it: which datasets were queried, with which filters, when, and with what coverage — so the search is reproducible, which is exactly what a Notified Body auditor wants to see.

Event counts are grouped by the FDA's reported event type (death, injury, malfunction). Name-based device matching is clearly labeled as indicative: only you can confirm whether a specific MAUDE report concerns your device. Recall records with implausible dates are excluded and the exclusion is disclosed. All data is from the official openFDA datasets and refreshed approximately monthly.