FDA Recall Terminated

00597001333 Femur CR PRC SURF HDN FEM CO-NID CML Rx, Sterile; 00597001334 Femur CR PRC SURF HDN FEM CO-NID CMR Rx, Sterile; 00597001433 Femur CR PRC SURF HDN FEM CO-NID DML Rx, Sterile; 00597001434 Femur CR PRC SURF HDN FEM CO-NID DMR Rx, Sterile; 00597001533 Femur CR PRC SURF HDN FEM CO-NID EML Rx, Sterile; 00597001534 Femur CR PRC SURF HDN FEM CO-NID EMR Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.

Recall: Z-2010-2013 · Initiated June 7, 2013

Recall

Recall Number
Z-2010-2013
Event Number
65413
Firm
Zimmer, Inc.
FEI Number
1000220733
Product Code
JWH
Status
Terminated
Root Cause
Labeling design
Initiated
June 7, 2013
Posted
August 22, 2013
Terminated
December 15, 2014
Address
345 E Main St, Warsaw, IN, 46580-2746

Description

00597001333 Femur CR PRC SURF HDN FEM CO-NID CML Rx, Sterile; 00597001334 Femur CR PRC SURF HDN FEM CO-NID CMR Rx, Sterile; 00597001433 Femur CR PRC SURF HDN FEM CO-NID DML Rx, Sterile; 00597001434 Femur CR PRC SURF HDN FEM CO-NID DMR Rx, Sterile; 00597001533 Femur CR PRC SURF HDN FEM CO-NID EML Rx, Sterile; 00597001534 Femur CR PRC SURF HDN FEM CO-NID EMR Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.

Reason

Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d

Action

Beginning June 7, 2013 all distributors will be sent Urgent Device Recall notices via electronic mail, with distributors, hospital risk managers with affected inventory also receiving the notice via courier. The Notice outlined the products recalled; the reason for the recall, possible adverse events related to the recall and directed consignees to locate, quarantine and count all affected product recording the information on the Inventory Return Form acknowledging that recall directions were followed. Further questions should be directed to 1-877-946-2761. Affected product is to be returned to Zimmer Product Service Department , 1777 West Center St. Warsaw, IN 46580 or internation return at Zimmer International Logistics GmbH, Max - Immelmann Allee 12 79427 Eschbach Germany. " Communications outside of the United States will occur approximately two weeks after the United States communications.

Distribution

Worldwide Distribution-USA (nationwide) and the country of Japan.

Quantity

192,355 all devices