FDA Recall Open, Classified

BD Pyxis MedStation ES, REF: 323, Medication Cabinet

Recall: Z-1927-2024 · Initiated April 15, 2024

Recall

Recall Number
Z-1927-2024
Event Number
94527
Firm
CareFusion 303, Inc.
FEI Number
2016493
Product Code
BRY
Status
Open, Classified
Root Cause
Component design/selection
Initiated
April 15, 2024
Posted
May 30, 2024
Address
10020 Pacific Mesa Blvd, San Diego, CA, 92121-4386

Description

BD Pyxis MedStation ES, REF: 323, Medication Cabinet

Reason

Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.

Action

On May 2, 2024, CareFusion, a subsidiary of Becton, Dickinson and company issued a "Urgent Medical Device Correction", to affected consignees. On September 30, 2025, BD issue a Urgent Medical Device Recall Expansion notification to affected consignees via FedEx. BD is asking consignees to take the following actions: 1. Affected products can continue to be used per existing Manuals and Guides. 2. In the event of an accidental fluid spill on the device or in a drawer, users should following instructions outlined within the customer notification. 3. Review the customer letter attachment which provides warnings, cautions and recommendations regarding fluid ingress. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material.

Distribution

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of South Africa, Kuwait, Qatar, Singapore, Bermuda, United Arab Emirates, Argentina, Australia, Belgium, Bahrain, Brazil, Bahamas, Canada, Chile, China, Germany, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Italy, Japan, South Korea, Lebanon, Morocco, Mexico, New Zealand, Oman, Portugal, Thailand, Taiwan.

Quantity

212,367 units