Ceiling Suspensions (Surgical Light and Flat Panel Suspensions); Catalog Numbers: 0682000863, 06820001009, 0682001010, 0682001076, 0682000864, 0682000857, 0682000858, 0682001077, 0682000865, 0682001011, 0682001078, 0682000866, 0682001012, 0682001080, 0682000867, 0682000959, 0682001079, 0682000850, 0682000851, 0682000852, 0682001269, 0682001567, 0682001569, 0682001570, 06820012691, 06820015671, 06820015691, & 0680015701. The ceiling suspensions are available in different configurations; they can hold monitors (flat panel suspensions), cameras (navigation suspensions arms) or surgical lights. The flat panel could have fixed or adjustable monitor yokes and may be single or dual arm configuration. Each ceiling suspension could hold a maximum of two surgical lights and either one camera or one monitor. There may be multiple ceiling/boom suspensions in an operating room. Intended for use as ceiling-mounted devices for supporting or positioning surgical lights, cameras, and/or monitors.
Recall
- Recall Number
- Z-1645-2010
- Event Number
- 55477
- Firm
- Stryker Communications Corp
- FEI Number
- 3002975342
- Product Code
- BRY
- Status
- Terminated
- Root Cause
- Employee error
- Initiated
- April 15, 2010
- Posted
- May 21, 2010
- Terminated
- January 24, 2011
- Address
- 1410 Lakeside Pkwy, Ste 100, Flower Mound, TX, 75028-4026
Description
Ceiling Suspensions (Surgical Light and Flat Panel Suspensions); Catalog Numbers: 0682000863, 06820001009, 0682001010, 0682001076, 0682000864, 0682000857, 0682000858, 0682001077, 0682000865, 0682001011, 0682001078, 0682000866, 0682001012, 0682001080, 0682000867, 0682000959, 0682001079, 0682000850, 0682000851, 0682000852, 0682001269, 0682001567, 0682001569, 0682001570, 06820012691, 06820015671, 06820015691, & 0680015701. The ceiling suspensions are available in different configurations; they can hold monitors (flat panel suspensions), cameras (navigation suspensions arms) or surgical lights. The flat panel could have fixed or adjustable monitor yokes and may be single or dual arm configuration. Each ceiling suspension could hold a maximum of two surgical lights and either one camera or one monitor. There may be multiple ceiling/boom suspensions in an operating room. Intended for use as ceiling-mounted devices for supporting or positioning surgical lights, cameras, and/or monitors.
Installation records are incomplete.
An Urgent: Medical Device Product Correction customer notification letter, dated 4/15/2010, has been sent out to all affected accounts via regular mail in the form of a certified, return request receipt letter. Accounts will receive a follow-up phone call from a Stryker Representative to schedule a re-inspection of equipment to see if it mets established critical specifications or not. The firm recommends that customers continue using all Stryker Communications products listed in the notice. Questions can be directed to Katie Blackwell at 972-410-7660 or via email at [email protected].
Nationwide Distribution