FDA Recall Terminated

iTotal PS Knee Replacement System, Model No. TPS-111-1111-010101, (ITOTAL PS IPOLY IMPLANT KIT LEFT) Product Usage: The iTotal CR Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.

Recall: Z-1628-2020 · Initiated February 3, 2020

Recall

Recall Number
Z-1628-2020
Event Number
85108
Firm
Conformis, Inc.
FEI Number
3009844603
Product Code
JWH
Status
Terminated
Root Cause
Mixed-up of materials/components
Initiated
February 3, 2020
Terminated
December 2, 2020
Address
600 Technology Park Dr, Billerica, MA, 01821-4154

Description

iTotal PS Knee Replacement System, Model No. TPS-111-1111-010101, (ITOTAL PS IPOLY IMPLANT KIT LEFT) Product Usage: The iTotal CR Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.

Reason

A limited number of tibial trays were labeled with an incorrect serial number and were delivered to the wrong customer accounts.

Action

Conformis Customer Service called three affected customers on February 4, 2020, to ask them to verify tibial tray serial numbers, and confirmed that the three kits in question had the incorrect tibial tray. These affected customers were asked to return the affected product to the firm. Following this action, Conformis Sales Representatives were instructed to verify that all products within their control, including those already placed in hospitals, that were manufactured between September 1, 2019, and February 3, 2020, had tibial tray serial numbers that matched the patient's serial number. No additional affected product was found in the field.

Distribution

US Nationwide distribution in the states of LA, NY, TN.

Quantity

2