FDA Recall Terminated

Universal Flat Panel Yoke, a component of the Stryker Flat Panel Arm. The yoke is intended to hold a flat panel monitor.

Recall: Z-1605-2010 · Initiated April 7, 2010

Recall

Recall Number
Z-1605-2010
Event Number
55498
Firm
Stryker Communications Corp
FEI Number
3002975342
Product Code
BRY
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
April 7, 2010
Posted
May 14, 2010
Terminated
March 30, 2011
Address
1410 Lakeside Pkwy, Ste 100, Flower Mound, TX, 75028-4026

Description

Universal Flat Panel Yoke, a component of the Stryker Flat Panel Arm. The yoke is intended to hold a flat panel monitor.

Reason

Some of the pitch shafts were not properly welded which results in the monitor becoming unlevel.

Action

An Urgent: Medical Device Recall letter dated April 19, 2010 was sent to customers. It was recommended that customers continue use of the product unless the monitor becomes unlevel. The customer was notified that a Stryker Representative would contact their facility to coordinate a service call to inspect the affected units. Questions are directed to the firm via phone at 972-410-7617 or via email at .

Distribution

Worldwide --US and Canada, Australia, China (Hong Kong), Latin America (Mexico, Chile, Ecuador, Costa Rica) .

Quantity

1498