FDA Recall
Terminated
Universal Flat Panel Yoke, a component of the Stryker Flat Panel Arm. The yoke is intended to hold a flat panel monitor.
Recall: Z-1605-2010
·
Initiated April 7, 2010
Recall
- Recall Number
- Z-1605-2010
- Event Number
- 55498
- Firm
- Stryker Communications Corp
- FEI Number
- 3002975342
- Product Code
- BRY
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- April 7, 2010
- Posted
- May 14, 2010
- Terminated
- March 30, 2011
- Address
- 1410 Lakeside Pkwy, Ste 100, Flower Mound, TX, 75028-4026
Description
Universal Flat Panel Yoke, a component of the Stryker Flat Panel Arm. The yoke is intended to hold a flat panel monitor.
Reason
Some of the pitch shafts were not properly welded which results in the monitor becoming unlevel.
Action
An Urgent: Medical Device Recall letter dated April 19, 2010 was sent to customers. It was recommended that customers continue use of the product unless the monitor becomes unlevel. The customer was notified that a Stryker Representative would contact their facility to coordinate a service call to inspect the affected units. Questions are directed to the firm via phone at 972-410-7617 or via email at .
Distribution
Worldwide --US and Canada, Australia, China (Hong Kong), Latin America (Mexico, Chile, Ecuador, Costa Rica) .
Quantity
1498