FDA Recall Terminated

Amico E-Cylinder Canister Holder, distributed by Stryker Communications, Flower Mound, TX 75028. Canister holder is used to hang E-Cylinders (pressurized gas canisters) from a boom. Product typically used in the surgical setting.

Recall: Z-1569-2009 · Initiated May 19, 2009

Recall

Recall Number
Z-1569-2009
Event Number
52099
Firm
Stryker Communications Corp
FEI Number
3002975342
Product Code
BRY
Status
Terminated
Root Cause
Device Design
Initiated
May 19, 2009
Posted
July 13, 2009
Terminated
April 26, 2010
Address
1410 Lakeside Pkwy, Ste 100, Flower Mound, TX, 75028-4026

Description

Amico E-Cylinder Canister Holder, distributed by Stryker Communications, Flower Mound, TX 75028. Canister holder is used to hang E-Cylinders (pressurized gas canisters) from a boom. Product typically used in the surgical setting.

Reason

Design flaw on the E-Cylinder canister holder may result in early fatigue and failure of the bracket screws, which could cause the cage to become detached from its mount and fall.

Action

An "Urgent: Device Recall" letter dated May 19, 2009 was issued to customers. Customers were advised to discontinue use of all affected E-cylinder holders. Consignees were asked to return the "Product Accountability Form" included with the recall letter, after which consignees will be provided with an improved cylinder holder at no cost. Direct questions to Stryker Communications at 1-972-410-7100.

Distribution

Nationwide and Canada.

Quantity

66 units.