FDA Recall Terminated

Depuy P.F.C. Modular Knee System, curved tibial insert (CVD); polyethylene tibial insert, Sz 2, 10.0 mm; Ref 98-4143.

Recall: Z-1334-05 · Initiated May 26, 2005

Recall

Recall Number
Z-1334-05
Event Number
32192
Firm
Depuy Orthopaedics, Inc.
FEI Number
1818910
Product Code
JWH
Status
Terminated
Root Cause
Other
Initiated
May 26, 2005
Posted
August 12, 2005
Terminated
December 2, 2005
Address
700 Orthopaedic Dr, Warsaw, IN, 46582-3994

Description

Depuy P.F.C. Modular Knee System, curved tibial insert (CVD); polyethylene tibial insert, Sz 2, 10.0 mm; Ref 98-4143.

Reason

An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.

Distribution

Nationwide, Argentina, Brazil, Canada, Caguas, Colombia, Ireland, Malaysia, Mexico, Panama, Peru, Singapore, South Africa, Sweden, Uruguay and Venezuela.

Quantity

10