Apollo Knee Tibial Baseplate/Knee Tibial Insert: CONG TIB INS SZ0/19MM APO CONG TIB INS SZ0/22MM APO CONG TIB INS SZ1/9MM APOL CONG TIB INS SZ1/11MM APO CONG TIB INS SZ1/19MM APO CONG TIB INS SZ2/13MM APO CONG TIB INS SZ2/19MM APO CONG TIB INS SZ3/9MM APOL CONG TIB INS SZ3/13MM APO
Recall
- Recall Number
- Z-1259-2014
- Event Number
- 67503
- Firm
- Zimmer, Inc.
- FEI Number
- 1000220733
- Product Code
- JWH
- Status
- Terminated
- Root Cause
- Storage
- Initiated
- February 25, 2014
- Posted
- March 20, 2014
- Terminated
- March 24, 2015
- Address
- 1800 W Center St, Warsaw, IN, 46580-2304
Description
Apollo Knee Tibial Baseplate/Knee Tibial Insert: CONG TIB INS SZ0/19MM APO CONG TIB INS SZ0/22MM APO CONG TIB INS SZ1/9MM APOL CONG TIB INS SZ1/11MM APO CONG TIB INS SZ1/19MM APO CONG TIB INS SZ2/13MM APO CONG TIB INS SZ2/19MM APO CONG TIB INS SZ3/9MM APOL CONG TIB INS SZ3/13MM APO
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
Beginning 2/25/2014 all distributors were notified via electronic mail. Hospital risk managers and distributors, surgeons with affected inventory will also be notified via courier or direct mail. The Urgent Lot Specific Recall Notice identified the issue and responsibilities. Distributors were asked to locate and remove the affected product in their territory, as well as identifying hospitals and surgeons who have previously implanted the affected product. Distributors were asked to return on-hand affected products to Zimmer and ensure all of their products are accounted for using the form provided in the letter and email it back to [email protected]. Hospital risk managers and surgeons were asked to locate affected product and return it to their Zimmer sales representative. Questions or concerns should be addressed to customer call center at 1-877-946-2761 between 8:00 am and 5:00pm EST Communications outside of the United States will occur approximately two weeks after the United States communications. 100 % effective checks will be done.
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
3638 units