FDA Recall Terminated

VIASYS Healthcare Disposable Bar Electrodes. Re-order No. 019-435600. Single Use only. Manufactured in the U.S.A. for VIASYS Healthcare, 5225 Verona Road, Bldg 2, Madison, WI USA 53711-4495..Each box contains 10 pouches of Bar Electrodes and each pouch contains 10 electrodes

Recall: Z-1254-06 · Initiated April 28, 2006

Recall

Recall Number
Z-1254-06
Event Number
35416
Firm
Nicolet Biomedical Div of Viasys Healthcare
FEI Number
3010611950
Product Code
GXZ
Status
Terminated
Root Cause
Other
Initiated
April 28, 2006
Posted
July 25, 2006
Terminated
December 27, 2006
Address
5225 Verona Rd, Madison, WI, 53711-4497

Description

VIASYS Healthcare Disposable Bar Electrodes. Re-order No. 019-435600. Single Use only. Manufactured in the U.S.A. for VIASYS Healthcare, 5225 Verona Road, Bldg 2, Madison, WI USA 53711-4495..Each box contains 10 pouches of Bar Electrodes and each pouch contains 10 electrodes

Reason

The Bar Electrodes do not meet performance specification for electrode impedance. A manufacturing defect was discovered in disposable bar electrodes, part number 019-435600, lot 604664, that has the potential to reduce the amplitude of a motor or sensory nerve action potential. These possible lower amplitudes will be measured only if the bar electrode is used as a bar.

Action

Communication on April 28, 2006 by phone, followed by Certified Mail, with return receipt to confirm receipt of information . The letter describes the issue and asks that affected product be returned. The customer notification allows the user the option of using the Bar Electrode as 2 separate electrodes, as described in the Instructions for Use.

Distribution

World wide including the states of AL, CA, CT, FL, IA, NH, NY, OH, TX. and countries of Canada, Japan, Poland

Quantity

53 boxes, 10 electrodes per pouch, 10 pouches per box