BD Pyxis Anesthesia ES System (PAS ES), REF: 327; BP Pyxis MedStation ES (Med ES Main), REF: 323; BD Pyxis MedStation ES Tower (Med ES Tower), REF: 352; all with BD Pyxis ES version 1.6.1 drawer firmware (version 1.10.2.5). Pyxis MedStation Enterprise Server (ES) System is an automated medication dispensing system supporting decentralized medication management in a pharmaceutical environment.

Recall: Z-1207-2023 · Initiated February 14, 2023

Recall

Recall Number
Z-1207-2023
Event Number
91341
Firm
CareFusion 303, Inc.
FEI Number
2016493
Product Code
BRY
Status
Open, Classified
Root Cause
Software change control
Initiated
February 14, 2023
Posted
March 17, 2023
Address
10020 Pacific Mesa Blvd, San Diego, CA, 92121-4386

Description

BD Pyxis Anesthesia ES System (PAS ES), REF: 327; BP Pyxis MedStation ES (Med ES Main), REF: 323; BD Pyxis MedStation ES Tower (Med ES Tower), REF: 352; all with BD Pyxis ES version 1.6.1 drawer firmware (version 1.10.2.5). Pyxis MedStation Enterprise Server (ES) System is an automated medication dispensing system supporting decentralized medication management in a pharmaceutical environment.

Reason

Automated dispensing cabinet drawer firmware may have inadvertently been downgraded to an earlier version of drawer firmware during new device installations or device replacement upgrades, which may lead to an increased frequency of drawer failures, which could increase the likelihood of patients experiencing delays in receiving medications.

Action

On 01/14/23, correction notices were mailed to directors of pharmacy informing them to take the following actions: 1) Ensure this letter and contents are forwarded and/or discussed with all appropriate personnel to ensure proper awareness of the issue. 2) If you are experiencing any issues, such as drawer failures, call BD technical support and refer to user guides for additional information and troubleshooting procedures. a. BD Pyxis MedStation ES user guide, refer to sections "Recover storage space" (Chapter 9, pg. 76) and "Emergency medication removal" (Appendix 11, pg. 119). See references 1 and 2 below for additional details. b. BD Pyxis PAS ES user guide, refer to sections "Emergency access procedures" (Appendix A, pg. 77) and "Storage spaces are not responsive" (Appendix B, pg. 81). See references 3 and 4 below for additional details. 3) Complete and return the Customer Response Form. The firm will take the following actions: 1) All customers who may have affected devices will be notified. 2) Upon receipt of completed response forms, the firm will coordinate the firmware upgrade with customers. Technical Support: 1-800-727-6102, Phone hours: 24 hours, 7 days a week

Distribution

Worldwide - US Nationwide distribution including in the states of NY, PA, OH, TX, FL, KY, IL, DE, AK, WI, KS, LA, WA, AZ, CO, MI, IN, TN, SD, AR, MO, CA, NV, MA, SC, VA, WV, CT, ME, NE, RI, GA, NC, IA, MN, AL, OK, MS, NM, NH, VT, WY, OR, ID, NJ, UT, DC, HI, MD, MT, ND and the country of Canada.

Quantity

7914