FDA Recall Terminated

Exprt Precision System: Revision Hip, Proximal Body with Bolt, Lateral Offset, Sterile R, djo surgical, REF: 495-00-065, 495-00-075, 495-00-085, 495-01-065, 495-01-075, 495-01-085

Recall: Z-1123-2019 · Initiated September 21, 2018

Recall

Recall Number
Z-1123-2019
Event Number
82379
FEI Number
1000116912
Product Code
KWZ
Status
Terminated
Root Cause
Other
Initiated
September 21, 2018
Terminated
September 16, 2020
Address
9800 Metric Blvd, Encore Medical, Lp, Austin, TX, 78758-5445

Description

Exprt Precision System: Revision Hip, Proximal Body with Bolt, Lateral Offset, Sterile R, djo surgical, REF: 495-00-065, 495-00-075, 495-00-085, 495-01-065, 495-01-075, 495-01-085

Reason

Complaints regarding loose proximal body bolts or bolts that were found through x-ray as sitting above the proximal body and not properly seated. Patient risks may include: pain, dissociation, and possible revision surgery

Action

On 09/21/18, Urgent Field Safety Notices, updated surgical techniques, and Acknowledgement and Receipt Forms were emailed to Surgical Sales Agents. The Surgical Technique provides instructions on use of the manual T-handle as the primary tightening method. It also provides instructions for use of the torque wrench to ensure that the readout for the torque wrench is at "0" before torque is applied. Sales agents were asked to affirm that they have read and understood the updated Surgical Technique information provided in the Field Safety Notice Bulletin. They were also asked to affirm that they disseminated this information to all employed and contracted sales reps in their organization. On 01/29/19, Urgent Field Safety Notices were emailed to Surgical Agents. This notice communicated the exchange process for physical product in inventory. The firm will notify Surgical Agents to make arrangements for the exchange of affected inventory with proximal bodies that are packaged with new capture bolts containing a polyethene plug. Surgical Agents with additional questions were encouraged to call 512-834-6255.

Distribution

US: TN, MN, TX, CO, IL, IN, OH, SC, MS, MI, RI, PA, ME, NY, WA, MO, AR, OK, LA, AL, FL, GA, UT, AZ, CA, NV,.

Quantity

22