Tool · Post-Market Surveillance

PSUR Report

The vigilance database section of your PSUR, generated in seconds: FDA MAUDE adverse event trends, recalls, and enforcement classifications for your device and similar devices — with the documented, reproducible methodology auditors ask for.

Period
from
to
Leave empty for the last 24 months

Reporting period 2024-072026-06 · FDA data refreshed ~monthly

Product Code: KWZ FDA class 2

Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer

View full classification →
Adverse events in period
326
+4% vs. prior period (312)
Deaths reported
0
Recalls in period
0
Class I enforcement
0

Adverse events per month

2024-07 – 2026-06
Loading...

Events by type

Period vs. prior period
Loading...
Event type
Period
Prior
Injury
294
278
Malfunction
32
34

Most reported coded problems

Top 15
Product problems
Count
Device Dislodged or Dislocated
147
Adverse Event Without Identified Device or Use Problem
112
Fracture
21
Unstable
16
Detachment of Device or Device Component
11
Naturally Worn
10
Difficult to Insert
8
Break
8
Scratched Material
6
Loosening of Implant Not Related to Bone-Ingrowth
5
Off-Label Use
4
Insufficient Device Problem Information
4
Osseointegration Problem
2
Mechanical Problem
2
Material Erosion
2
Patient problems
Count
Joint Dislocation
152
Insufficient Information
43
Unspecified Infection
36
Joint Laxity
33
No Clinical Signs, Symptoms or Conditions
32
Pain
21
Medical device site infection
9
Ambulation or Postural Difficulties
8
Fall
6
Failure of Implant
6
Bone Fracture(s)
5
Osteolysis
4
Aseptic loosening
4
Sepsis
3
Scar Tissue
3

Recalls in period

0 total

No recalls recorded for these product codes in the reporting period.

Adverse events by year

Loading...

Methodology & data provenance

For your PSUR appendix

Search protocol: FDA MAUDE adverse event reports, FDA device recalls, and FDA enforcement reports (openFDA datasets) were queried for product code KWZ, reporting period 2024-07 to 2026-06 (prior comparison period 2022-07 to 2024-06).

Data coverage: adverse event data for these product codes extends through 2026-05. Source openFDA export dated 2026-06-30. Recall records with missing or implausible initiation dates (0 for these codes) are excluded from period counts.

Generated: 2026-07-05 23:49 UTC by BEUDAMED (beudamed.com/fda/psur-report). Dataset sizes: 25,039,198 adverse events, 58,607 recalls, 39,365 enforcement reports.

PSUR database searches, explained

Under the EU MDR, manufacturers must produce a Periodic Safety Update Report (PSUR) (Class IIa and above; a PMS report for Class I) on a defined schedule. A required part of that report is a survey of publicly available vigilance data — most importantly the FDA's MAUDE adverse event database and FDA recall records — covering both your own device and similar devices on the market (the "state of the art").

Doing this by hand means repeated MAUDE queries, manual export, deduplication, and trending — hours of work per report, repeated every reporting period. This tool runs the same search protocol in seconds and documents it: which datasets were queried, with which filters, when, and with what coverage — so the search is reproducible, which is exactly what a Notified Body auditor wants to see.

Event counts are grouped by the FDA's reported event type (death, injury, malfunction). Name-based device matching is clearly labeled as indicative: only you can confirm whether a specific MAUDE report concerns your device. Recall records with implausible dates are excluded and the exclusion is disclosed. All data is from the official openFDA datasets and refreshed approximately monthly.