FDA Recall Terminated

B5-2Q BALLOON CATHETER, Model No. B5-2Q - Product Usage: Used with endoscopes to inject contrast medium or other medical fluid into the bile duct, pancreatic duct, and respiratory organs.

Recall: Z-0961-2021 · Initiated January 4, 2021

Recall

Recall Number
Z-0961-2021
Event Number
87074
Firm
Aomori Olympus Co., Ltd. 2 Chome 248-1 Okkonoki Kuroishi Japan
FEI Number
3003995201
Product Code
FGE
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
January 4, 2021
Terminated
March 11, 2024

Description

B5-2Q BALLOON CATHETER, Model No. B5-2Q - Product Usage: Used with endoscopes to inject contrast medium or other medical fluid into the bile duct, pancreatic duct, and respiratory organs.

Reason

Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.

Action

On January 4, 2021, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were informed that the identified list of Endo-therapy devices may have a defective seal which may compromise the sterility of the product. Customers were asked to take the following actions: 1. Immediately assess any product you have in stock to identify endotherapy products with affected lot numbers listed in the recall letter. Cease use of product and quarantine any affected product. 2. Call your Olympus customer service representative at 1-888-524-7266 to obtain a Returned Goods Authorization so that you may return the product with no charge to you. Olympus will issue a credit or replacement to your facility for any returned product. 3. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number Recall-0391 and provide your contact information as indicated in the portal. 4. If you have distributed these devices outside your facility, please notify your customers of this corrective action immediately. Your notification to your customers may be enhanced by including a copy of this recall notification letter.

Distribution

US Nationwide distribution.

Quantity

7,400,000 (global); 260,395 (US)