FDA Recall Open, Classified

Becker External Drainage and Monitoring System: 24146 EDMS 24146 BECKER NO Y-SITE 25120 EDMS II 25120 BECKER W/1-WAY VALVE 26040 KIT 26040 BECKER EDMS II 27609 BECKER 27609 W/NEEDLESLESS INJ. SITE 27670 BECKER 27670 STOP BELOW DC NDL-LS IN 27672 BECKER 27672 ONEWAY VLV Y-SITE PT LN 27702 EDMS 27702 EURO BECKER II GR PT LN 27761 BECKER 27761 STOP BCNDLESS PATLN 24 27767 BECKER 27767 NO STPCOCK ON PTLN 63 27779 EDMS 27779 BECKER BC NDLESS BOND CONN 27811 EDMS 27811 BECKER GRN PL VCATH 27931 EDMS 27931 BECKER GS TUBING SMARTSITES 46128 EDMS 46128 BECKER II BLUE PT LN 46129 EDMS 46129 EURO BECKER II BLUE PT LN The Becker External Drainage and Monitoring System (Becker EDMS) is a closed system for: 1. Draining cerebrospinal fluid (CSF) from the lateral ventricles of the brain or the lumbar subarachnoid space. 2. Monitoring CSF pressure and flow rate from the lateral ventricles of the brain and the lumbar subarachnoid space.

Recall: Z-0661-2025 · Initiated November 1, 2024

Recall

Recall Number
Z-0661-2025
Event Number
95725
Firm
Medtronic Neurosurgery
FEI Number
3015531529
Product Code
GWM
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
November 1, 2024
Posted
December 19, 2024
Address
4620 North Beach Street, Fort Worth, TX, 76137

Description

Becker External Drainage and Monitoring System: 24146 EDMS 24146 BECKER NO Y-SITE 25120 EDMS II 25120 BECKER W/1-WAY VALVE 26040 KIT 26040 BECKER EDMS II 27609 BECKER 27609 W/NEEDLESLESS INJ. SITE 27670 BECKER 27670 STOP BELOW DC NDL-LS IN 27672 BECKER 27672 ONEWAY VLV Y-SITE PT LN 27702 EDMS 27702 EURO BECKER II GR PT LN 27761 BECKER 27761 STOP BCNDLESS PATLN 24 27767 BECKER 27767 NO STPCOCK ON PTLN 63 27779 EDMS 27779 BECKER BC NDLESS BOND CONN 27811 EDMS 27811 BECKER GRN PL VCATH 27931 EDMS 27931 BECKER GS TUBING SMARTSITES 46128 EDMS 46128 BECKER II BLUE PT LN 46129 EDMS 46129 EURO BECKER II BLUE PT LN The Becker External Drainage and Monitoring System (Becker EDMS) is a closed system for: 1. Draining cerebrospinal fluid (CSF) from the lateral ventricles of the brain or the lumbar subarachnoid space. 2. Monitoring CSF pressure and flow rate from the lateral ventricles of the brain and the lumbar subarachnoid space.

Reason

Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).

Action

On 11/01/2024, the firm sent an URGENT: MEDICAL DEVICE CORRECTION Letter via US mail to customers informing them that through customer complaints, the firm is aware of potential cracking and leaking in the "stopcock" components on the Becker and Exacta External Drainage and Monitoring Systems. Customer are instructed to: Recommended Mitigations: -Prior to use, inspect all stopcocks and connections to ensure that connections are secure and there are no visible cracks in the stopcocks. If cracks or leaks are identified, do not use return the device to Medtronic. -The system must be pre-filled with sterile isotonic saline solution prior to connecting to the patient and all connections checked to ensure that fittings (connections) are tight and leak-free. -All connections should be finger tightened. Over tightening can cause cracks and leaks to occur. -After cleaning with alcohol, or a disinfectant containing alcohol, allow to air dry completely prior to connecting the system. Patient Management Recommendations: -If system develops cracks/leaks during use, replace using sterile technique and return the damaged system to Medtronic. If leaks occur with the Patient Line Stopcock, consider application of a hemostat or other clamp to the proximal patient line while coordinating a replacement if occlusion of the patient line does not create a risk to the patient. -In the product labeling, all patients with EDMS should be monitored for evidence of infection. If cracks or leaks, the system should be replaced, and the patient should continue to be monitored for evidence of infection. If customer chooses to return unused product contact Medtronic Customer Service via email: [email protected], or Phone: 1-800-874-5797 Option 1, and mention FA1452, to initiate return process. For questions-contact Medtronic Rep/Medtronic Technical Support (US) at 1-888-826-5603and reference FA1452.

Distribution

Worldwide distribution. US nationwide including Puerto Rico; Algeria, Argentina, Australia, Austria, Bahrain, Barbados, Bosnia And Herzegovina, Brunei Darussalam, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, French Guiana, French Polynesia, Germany, Ghana, Greece, Guadeloupe, Honduras, Hong Kong, Iceland, Indonesia, Israel, Italy, Jamaica, Jordan, Kazakhstan, Kenya, Kuwait, Lao People s Democratic Republic, Latvia, Lebanon, Libya, Malaysia, Malta, Mauritania, Morocco, Myanmar, Namibia, Netherlands, New Zealand, Nicaragua, Norway, Oman, Panama, Peru, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Republic of Tanzania, Thailand, Trinidad And Tobago, Tunisia, Turkmenistan, United Arab Emirates, United Kingdom, Uruguay, Vietnam, and Yemen.

Quantity

65,810 units