Reporting period 2024-072026-06 · FDA data refreshed ~monthly

Product Code: GWM FDA class 2

Device, Monitoring, Intracranial Pressure

View full classification →
Adverse events in period
852
-35% vs. prior period (1,309)
Deaths reported
4
Recalls in period
7
Class I enforcement
2

Adverse events per month

2024-07 – 2026-06
Loading...

Events by type

Period vs. prior period
Loading...
Event type
Period
Prior
Death
4
8
Injury
224
280
Malfunction
623
1,021
Other
1
0

Most reported coded problems

Top 15
Product problems
Count
Fluid/Blood Leak
217
Break
165
Fracture
151
Incorrect, Inadequate or Imprecise Result or Readings
98
Unexpected Shutdown
60
Incorrect Measurement
56
Disconnection
46
Device Sensing Problem
21
Output Problem
20
Display or Visual Feedback Problem
20
Insufficient Device Problem Information
18
Signal Artifact/Noise
16
Device Dislodged or Dislocated
15
No Device Output
14
Infusion or Flow Problem
14
Patient problems
Count
No Clinical Signs, Symptoms or Conditions
633
Insufficient Information
165
Cerebrospinal Fluid Leakage
18
Unspecified Infection
13
Cerebral Edema
4
Foreign Body In Patient
3
Device Embedded In Tissue or Plaque
3
Bacterial Infection
3
Visual Disturbances
2
Pneumocephalus
2
Vomiting
1
Vessel Or Plaque, Device Embedded In
1
Skin Inflammation/ Irritation
1
Pain
1
Nausea
1

Recalls in period

7 total
FDA enforcement classification: Class I: 2 Class II: 27
Date
Recalling firm
Status
2025-12-12
Open, Classified
2025-12-12
Open, Classified
2025-11-17
Open, Classified
2024-11-01
Open, Classified
2024-11-01
Open, Classified
2024-09-24
Open, Classified
2024-09-24
Open, Classified

Adverse events by year

Loading...

Methodology & data provenance

For your PSUR appendix

Search protocol: FDA MAUDE adverse event reports, FDA device recalls, and FDA enforcement reports (openFDA datasets) were queried for product code GWM, reporting period 2024-07 to 2026-06 (prior comparison period 2022-07 to 2024-06).

Data coverage: adverse event data for these product codes extends through 2026-05. Source openFDA export dated 2026-06-30. Recall records with missing or implausible initiation dates (0 for these codes) are excluded from period counts.

Generated: 2026-07-05 23:01 UTC by BEUDAMED (beudamed.com/fda/psur-report). Dataset sizes: 25,039,198 adverse events, 58,607 recalls, 39,365 enforcement reports.

PSUR database searches, explained

Under the EU MDR, manufacturers must produce a Periodic Safety Update Report (PSUR) (Class IIa and above; a PMS report for Class I) on a defined schedule. A required part of that report is a survey of publicly available vigilance data — most importantly the FDA's MAUDE adverse event database and FDA recall records — covering both your own device and similar devices on the market (the "state of the art").

Doing this by hand means repeated MAUDE queries, manual export, deduplication, and trending — hours of work per report, repeated every reporting period. This tool runs the same search protocol in seconds and documents it: which datasets were queried, with which filters, when, and with what coverage — so the search is reproducible, which is exactly what a Notified Body auditor wants to see.

Event counts are grouped by the FDA's reported event type (death, injury, malfunction). Name-based device matching is clearly labeled as indicative: only you can confirm whether a specific MAUDE report concerns your device. Recall records with implausible dates are excluded and the exclusion is disclosed. All data is from the official openFDA datasets and refreshed approximately monthly.