FDA Recall Terminated

LEGION(TM) COCR CONSTRAINED FEMORAL COMPONENT, SIZE 5, RIGHT, REF 71426005, QTY 1, Smith & Nephew, knee prosthesis Usage:orthopedic.

Recall: Z-0590-2015 · Initiated November 4, 2014

Recall

Recall Number
Z-0590-2015
Event Number
69759
Firm
Smith & Nephew, Inc.
FEI Number
1020279
Product Code
JWH
Status
Terminated
Root Cause
Process control
Initiated
November 4, 2014
Posted
December 12, 2014
Terminated
August 22, 2019
Address
1450 E Brooks Rd, Memphis, TN, 38116-1804

Description

LEGION(TM) COCR CONSTRAINED FEMORAL COMPONENT, SIZE 5, RIGHT, REF 71426005, QTY 1, Smith & Nephew, knee prosthesis Usage:orthopedic.

Reason

The stem taper on this device is undersized; the gauge point is not deep enough.

Action

The firm, Smith & Nephew, Inc., sent an "Urgent-Product Recall 1st Notification-Urgent" letter dated 11/4/ 2014 to its customers via E-mail on 11/4/2014. International distributors were notified on 11/4/2014 by Email. Non-responding consignees will be contacted on 11/25/2014 by original method. The letter described the product, problem, and actions to be taken. The customers were instructed to inspect your inventory and locate any unused devices and immediately quarantine; complete and return the Inventory Return Certification Form via fax to: +1-901-566-7975, even if you do not have product to return; contact Smith & Nephew's Safety Affairs Department via email at [email protected] or fax +1-901-566-7975 to obtain return authorization (RA) number; and return affected product to Smith & Nephew, Attn: Global Field Actions, 3303 E Holmes Road, Memphis, TN 38118. If you have any questions, please contact the Group Director, Quality Assurance +1-901-399-1970.

Distribution

Worldwide Distribution: US (nationwide) including states of: MI, MN, MO, OH, PA, IA, and TN; and Internationally to: SWITZERLAND; CANADA, COLOMBIA, GERMANY, Dubai, (UAE), THAILAND; SOUTH AFRICA, SOUTH KOREA; and SINGAPORE.

Quantity

45 units