Smith & Nephew GENESIS(R) II/LEGION(TM) PATELLAR COMPONENT, RESURFACING ROUND, 32MM STANDARD, FOR CEMENTED USE ONLY WITH JOURNEY PEG, STERILE EO, Rx only, REF Number: 71934490 knee implant.

Recall: Z-0573-2016 · Initiated November 9, 2015

Recall

Recall Number
Z-0573-2016
Event Number
72866
Firm
Smith & Nephew, Inc.
FEI Number
1020279
Product Code
JWH
Status
Terminated
Root Cause
Process design
Initiated
November 9, 2015
Posted
January 4, 2016
Terminated
March 17, 2016
Address
1450 E Brooks Rd, Memphis, TN, 38116-1804

Description

Smith & Nephew GENESIS(R) II/LEGION(TM) PATELLAR COMPONENT, RESURFACING ROUND, 32MM STANDARD, FOR CEMENTED USE ONLY WITH JOURNEY PEG, STERILE EO, Rx only, REF Number: 71934490 knee implant.

Reason

One lot of 35 mm patella implants is labeled as 32 mm. These implants have pegs which are spaced proportionally; therefore the mismatch prevents the pegs from properly aligning with pre-drilled holes during implantation.

Action

The firm, Smith & Nephew Inc, notified the distributors by Fed- Ex on 10/20/2015 and sent an "Urgent - Product Recall 1st Notification - Urgent R-2015-22" letter dated 11/9/2015 to their customers. The letter described the product, problem, and actions to be taken. The customers were instructed to inspect your inventory and locate any unused devices, quarantine them immediately, complete and return the Inventory Return Certification Form by fax to: +1-901-566-7975 or email to: [email protected], even if you do not have the product, contact Smith & Nephew's Field Actions Department via email or fax (above) to obtain a return authorization (RA) number, and return any affected product to Smith & Nephew, Attn: Global Field Actions, Global Distribution Center, 3303 E Holmes Road, Memphis, TN 38118. If you have any questions, call Director, Regulatory Affairs 901-399-5520 or email: [email protected].

Distribution

US Distribution to states of:FL, MO, IN and TX.

Quantity

20 units