JOURNEY II BCS CONSTRAINED ARTICULAR INSERT Size 5-6, 10 MM Left, REF 74029262; knee prosthesis
Recall
- Recall Number
- Z-0158-2024
- Event Number
- 93156
- Firm
- Smith & Nephew, Inc.
- FEI Number
- 1020279
- Product Code
- JWH
- Status
- Open, Classified
- Root Cause
- Component change control
- Initiated
- September 27, 2023
- Posted
- October 24, 2023
- Address
- 1450 E Brooks Rd, Memphis, TN, 38116-1804
Description
JOURNEY II BCS CONSTRAINED ARTICULAR INSERT Size 5-6, 10 MM Left, REF 74029262; knee prosthesis
The JRNY II BCS XLPE ART ISRT SZ 5-6 LT 10MM was laser etched, labeled and packaged as a JRNY II BCS CNSTRD ART ISRT 5-6 LT 10MM. And also, a JRNY II BCS CNSTRD ART ISRT 5-6 LT 10MM was laser etched, labeled and packaged as a JRNY II BCS XLPE ART ISRT SZ 5-6 LT 10MM.
Smith & Nephew, Inc. issued an Urgent Medical Device Recall Notice to its consignees on 09/27/2023 via email and overnight mail. The notice explain the issue, potential risk, and requested the following actions: Inspect inventory, locate, and quarantine the affected devices. Sales representatives, district offices, or distributors were directed to notify their customers The firm is seeking return of unused products.
Worldwide distribution - US Nationwide and the country of Japan.
29 units