Sidus" Head (Not distributed in the United States)
Recall
- Recall Number
- Z-0079-2014
- Event Number
- 65799
- Firm
- Zimmer, Inc.
- FEI Number
- 1000220733
- Product Code
- JWH
- Status
- Terminated
- Root Cause
- Package design/selection
- Initiated
- August 30, 2013
- Posted
- October 25, 2013
- Terminated
- September 25, 2014
- Address
- 345 E Main St, Warsaw, IN, 46580-2746
Description
Sidus" Head (Not distributed in the United States)
The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.
August 30, 2013 Urgent Recall Notices were sent to Distributors, hospital risk managers and surgeons identifying the issue and directing consignees to make all affected personnel aware of the contents of the notification. The notice identified the products involved, the issue, possible adverse events associated with the issue and recommended that implants found with LDPE bag adhesion not use the product and return it to Zimmer sales representatives. Questions or concerns should be addressed to Zimmer customer call center at 1-877-946-2761 between the hours of 8 a.m. and 5 p.m. EST.
Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Germany, Guatemala, Honduras, Hong Kong, India, Israel, Jamaica, Japan, Korea, Malaysia, Mexico, New Zealand, Nicaragua, Panama, Peru, Singapore, Taiwan, Thailand, and Venezula.
0 in USA