FDA Adverse Event Injury Summary report: N

GMK-SPHERE FEMORAL COMPONENT CEMENTED SIZE 3+ R

MDR report key: 9880470 · Received March 25, 2020

Report

Report Number
3005180920-2020-00177
Event Type
Injury
Date Received
March 25, 2020
Date of Event
February 24, 2020
Report Date
March 25, 2020
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JWH
UDI-DI
07630030862380
PMA / PMN Number
K140826
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 05 MARCH 2020: LOT 1810113: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 07-FEB-2019. EXPIRATION DATE: 2024-01-23. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. ADDITIONAL DEVICES INVOLVED BATCH REVIEWS PERFORMED ON 05 MARCH 2020: GMK-SPHERE 02.12.0210FR TIBIAL INSERT FIXED SPHERE FLEX SIZE 2/10 MM R (K121416) LOT 186055: (B)(4)ITEMS MANUFACTURED AND RELEASED ON 29-NOV-2018. EXPIRATION DATE: 2023-11-19. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. GMK-SPHERE 02.07.1202R TIBIAL TRAY FIXED CEMENTED SIZE 2 R (K090988) LOT 171148: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 28-JUNE-2017. EXPIRATION DATE: 2022-06-12. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

THE PATIENT CAME IN REPORTING STIFFNESS AND PAIN IN THE KNEE. THE CAUSE OF THE KNEE STIFFNESS AND PAIN IS PATELLA BAJA. 7 MONTHS AFTER PRIMARY SURGERY THE SURGEON REVISED SUCCESSFULLY THE KNEE SWAPPING THE FEMORAL AND TIBIAL COMPONENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
344483 GMK-SPHERE FEMORAL COMPONENT CEMENTED SIZE 3+ R KNEE CEMENTED FEMUR JWH MEDACTA INTERNATIONAL SA 02.12.0023R 1810113 07630030862380

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention