UNKNOWN NEXGEN TIBIAL TRAY
Report
- Report Number
- 0001822565-2019-04902
- Event Type
- Malfunction
- Date Received
- December 3, 2019
- Date of Event
- March 12, 2019
- Report Date
- November 4, 2020
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- JWH
- PMA / PMN Number
- N/A
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- PHYSICIAN
Narratives
THIS FOLLOW-UP IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. VISUAL EXAMINATION OF THE RETURNED PRODUCT/ PROVIDED PICTURES SHOWS THAT THE DISTAL SURFACE TO EXHIBIT DAMAGE (NICKED GOUGED) AND THE DOVE TAIL FEATURES TO BE COMPRESSED AND FLARED OUT. DAMAGE TO THE DOVETAIL FEATURE CONFIRMS THE IMPLANT WOULD NOT BE ABLE TO SEAT. MEDICAL RECORDS WERE NOT PROVIDED. DEVICE HISTORY RECORD (DHR) REVIEW WAS UNABLE TO BE PERFORMED AS THE PART AND LOT NUMBER OF THE DEVICE INVOLVED IN THE EVENT IS UNKNOWN. DAMAGE TO THE ARTICULAR SURFACE CAN OCCUR IF THE ARTICULAR SURFACE IS NOT CORRECTLY PLACED AND ORIENTED BEFORE PUSHING IT USING THE INSERTER INSTRUMENT. DETAILED INSTRUCTIONS FOR INSERTING THE ARTICULAR SURFACE ARE PROVIDED IN LPS-FLEX FIXED BEARING KNEE SURGICAL TECHNIQUE (97-5964-102-00-REV0616) PG. 15-17. THE ROOT CAUSE IS ATTRIBUTED TO USE ERROR. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
(B)(4). MEDICAL PRODUCT: ARTICULAR SURFACE SIZE CD 12 MM HEIGHT "USE WITH PLATE 1. ITEM# 00596402212 LOT# 63631104. THE REPORTED EVENT CANNOT BE CONFIRMED. NO PRODUCT WAS RETURNED OR PICTURES PROVIDED; THEREFORE, VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. MEDICAL RECORDS WERE NOT PROVIDED. DEVICE HISTORY RECORD (DHR) REVIEW WAS UNABLE TO BE PERFORMED AS THE PART AND LOT NUMBER OF THE DEVICE INVOLVED IN THE EVENT IS UNKNOWN. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT THE ARTICULAR SURFACE WOULD NOT LOCK ONTO THE TIBIAL PLATE. THERE IS NO ADDITIONAL INFORMATION AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1201760 | UNKNOWN NEXGEN TIBIAL TRAY | PROSTHESIS, KNEE | JWH | ZIMMER BIOMET, INC. | N/A | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR |