FDA Adverse Event Injury Summary report: N

VGXP XP E1 TIB BRG

MDR report key: 8796689 · Received July 16, 2019

Report

Report Number
0001825034-2019-03045
Event Type
Injury
Date Received
July 16, 2019
Report Date
December 13, 2019
Manufacturer
ZIMMER BIOMET, INC.
Product Code
JWH
PMA / PMN Number
K132873
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY UPDATED AND ADDITIONAL INFORMATION. COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS NOT CONFIRMED. VISUAL EXAMINATION OF THE RETURNED BEARING FOUND DEEP GOUGES AND LIGHT SCRATCHES. DHR WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. ROOT CAUSE WAS UNABLE TO BE DETERMINED. A SUMMARY OF THE INVESTIGATION HAS BEEN SENT TO THE COMPLAINANT. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT MEDICAL PRODUCTS: VGXP XP E1 TIB BRG; P/N: 195815, L/N: 822870, KNE-VANGUARD-LOCKING BAR-UNK; P/N: UNK, L/N: UNK. PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034 - 2019 - 03045, 0001825034 - 2019 - 03046.

Description of Event or Problem · 1

IT WAS REPORTED PATIENT HAD A REVISION ON AN UNKNOWN DATE DUE TO THE LOCKING BAR BACKING OUT OF TIBIA AND THE LATERAL BEARINGS DISLOCATING. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
588737 VGXP XP E1 TIB BRG PROSTHESIS, KNEE JWH ZIMMER BIOMET, INC. N/A 822870

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R