FDA Adverse Event Injury Summary report: N

AMICUS SEPARATOR SYSTEM

MDR report key: 8166374 · Received December 14, 2018

Report

Report Number
3004548776-2018-00004
Event Type
Injury
Date Received
December 14, 2018
Date of Event
October 11, 2018
Report Date
November 28, 2018
Manufacturer
FENWAL INTERNATIONAL INC.
Product Code
GKT
UDI-DI
04086000101325
PMA / PMN Number
BK960005
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 0

THE FOLLOWING INFORMATION WAS SENT TO THE FDA ON MARCH 14, 2019 IN RESPONSE TO A LETTER RECEIVED ON JANUARY 28, 2019 ASKING FOR ADDITIONAL INFORMATION REGARDING THIS REPORTED INCIDENT. FRESENIUS KABI IS THE MANUFACTURER OF THE AMICUS SEPARATOR SYSTEM. THE AMERICAN RED CROSS (ARC) USES THE MEDICAL DEVICE TO COLLECT PLATELETS FROM A DONOR AND THEN FURTHER MANUFACTURES THE PLATELET PRODUCT. THE COLLECTED PLATELET PRODUCT IS THEN DISTRIBUTED TO ARC'S CUSTOMERS SUCH AS HOSPITALS AND\OR OTHER FACILITIES FOR PATIENTS NEEDING PLATELET TRANSFUSIONS. THE INCIDENT REPORTED IN THIS MDR WAS REPORTED TO US BY ARC BUT PERTAINS TO TWO PATIENTS WHOM EXPERIENCED AN ADVERSE REACTION AT THE HOSPITAL/FACILITY THAT RECEIVED THE PLATELETS (FROM THE SAME DONATION) FROM AMERICAN RED CROSS (NOT DIRECTLY FROM FRESENIUS KABI). FRESENIUS KABI DOES NOT HAVE THE INFUSED PATIENT'S INFORMATION. COMPLAINT (B)(4) WAS RECEIVED AGAINST MATERIAL CODE 4R2340, LOT NUMBER FA18C26074. THE CUSTOMER STATES THAT A FACILITY USING THEIR COLLECTED PLATELET PRODUCT EXPERIENCED AN ADVERSE REACTION POST PLATELET TRANSFUSION (ASSUMED TO BE DUE TO BACTERIAL CONTAMINATION). NEITHER A SAMPLE NOR PICTURE WERE AVAILABLE FOR FURTHER EVALUATION. COMPLAINT (B)(4) INVESTIGATION RESULTS: REVIEWED LOT MANUFACTURING RECORDS FOR ANY DOCUMENTED EXCEPTIONS ON THE FINISHED GOOD LOT AND ASSOCIATED SOLUTION BATCHES. NO EXCEPTIONS WERE FOUND. REVIEWED MANUFACTURING RECORDS FOR RESULTS OF IN-PROCESS AND FINAL RELEASE QUALITY TESTS. RESULTS WERE SATISFACTORY. THE BACTERIA REPORTED WAS NOT IDENTIFIED IN THE HAINA, DOMINICAN REPUBLIC MICROBIAL FLORA. PERFORMED A REVIEW ON COMPLAINTS RECEIVED FOR AFFECTED LOT AND OTHER AMICUS KIT BATCHES THAT USED THE SAME SOLUTION BATCHES. NO COMPLAINTS HAVE BEEN REPORTED FOR ADVERSE REACTIONS OR BACTERIAL CONTAMINATION ON ANY OF THE LOTS REVIEWED. A REVIEW OF CAPAS WAS CONDUCTED AND THERE ARE NO CAPAS RELATED TO THE DEFICIENCY LISTED IN THIS COMPLAINT. CONCLUSION: THE ABOVE INVESTIGATION FOR THIS LOT DOES NOT REVEAL ANY PROCESS, PRODUCT OR MANUFACTURING NONCONFORMITY THAT WOULD CONTRIBUTE TO BACTERIAL CONTAMINATION. THE LOT MEETS ALL THE RELEASE REQUIREMENTS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT A FACILITY USING THEIR COLLECTED PLATELET PRODUCT REPORTED 2 PATIENTS EXPERIENCED AN ADVERSE REACTION POST PLATELET TRANSFUSION. THE PATIENTS WERE INFUSED WITH PLATELETS COLLECTED DURING THE SAME DONATION. THE PLATELET PRODUCT WAS COLLECTED ON (B)(6) 2018. THE COLLECTED PLATELETS WERE TREATED WITH PAS. BOTH PATIENTS WERE TRANSFUSED ON (B)(6) 2018. POST PLATELET TRANSFUSION, THE PLATELET PRODUCT TESTED POSITIVE FOR ACINETOBACTER BAUMANNII. THE FOLLOWING INFORMATION HAS BEEN REPORTED TO US BY THE (B)(6) REGARDING THE CDC INVESTIGATION: A. BAUMANNII WAS ISOLATED AND THE FOLLOWING CDC INVESTIGATION RESULTS ARE CONSIDERED RELEVANT TO THE INVESTIGATION: THE ISOLATE IN THIS INSTANCE IS SIMILAR BUT DISTINCT FROM THE OTHER 2 RECENTLY REPORTED CASES IN 2018. A BATCH TRACEABILITY IS BEING PERFORMED FOR THE SOLUTIONS USED IN AMICUS KIT PRODUCT CODE 4R2340, BATCH FA18C26074. THERE ARE NO COMMON PRE-ATTACHED SOLUTION SUBASSEMBLY LOT NUMBERS BETWEEN THE TWO AMICUS KIT LOTS INVOLVED IN THREE RECENTLY REPORTED CASES IN 2018. THERE ARE NO RELATED COMPLAINTS AGAINST OTHER AMICUS KIT BATCHES THAT USED THE SAME SOLUTION BATCHES. BOTH PRODUCTION PLANTS INVOLVED IN THE MANUFACTURE OF AMICUS KITS WERE TESTED FOR THE PRESENT BACTERIA. NEITHER PLANT TESTED POSITIVE FOR ACINETOBACTER BAUMANNII COMPLEX. INVESTIGATION RESULTS THUS FAR: THERE IS NO EVIDENCE IMPLICATING OUR PRODUCTS AS A ROOT CAUSE FOR THE ACINETOBACTER CONTAMINATION INCIDENCES AS SHOWN IN THE FOLLOWING ANALYSES. A CAPA INVESTIGATION HAS BEEN OPENED AND THE NEXT STEPS WERE DISCUSSED WITH THE FDA: 3RD PARTY MICROBIOLOGIC TESTS AT BOTH PRODUCTION UNITS HAVE BEEN INITIATED. TESTING ON THE ARC ISOLATES OF ACINETOBACTER HAVE BEEN INITIATED TO EVALUATE RESISTANCE TO MOIST HEAT AND RADIATION. REVIEWING MANUFACTURING CAPA, ISSUES AND NONCONFORMITIES TO DETERMINE IF ANY HAVE THE ABILITY TO IMPACT INTEGRITY AND/OR STERILITY OF THE KITS. INVESTIGATIONS FROM FDA/CDC HAVE BEEN REQUESTED TO CONFIRM DATA ALONG WITH A REQUEST FOR DNA SEQUENCE DATA, NOT AVAILABLE TO THE PUBLIC AT THIS TIME. WITH THE INVESTIGATION IN PROCESS, A FOLLOW-UP REPORT WILL BE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1005636 AMICUS SEPARATOR SYSTEM AMICUS APHERESIS KIT GKT FENWAL INTERNATIONAL INC. N/A FA18C26074 04086000101325

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening