FDA Adverse Event Malfunction Summary report: N

BMT 360 2.5MM OFFSET ADAPTER

MDR report key: 8052745 · Received November 8, 2018

Report

Report Number
0001825034-2018-10283
Event Type
Malfunction
Date Received
November 8, 2018
Date of Event
November 1, 2018
Report Date
August 29, 2019
Manufacturer
ZIMMER BIOMET, INC.
Product Code
JWH
PMA / PMN Number
K093293
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE COMPLAINT IS CONSIDERED CONFIRMED AS VISUAL INSPECTION OF THE SET SCREW SHOWS THE INSIDE OF THE HEX HAS BEEN ROUNDED OFF. THE TIBIAL TRAY, CRUCIATE WING, OFFSET ADAPTOR AND SET SCREW ARE STILL ASSEMBLED. DEVICE HISTORY RECORD (DHR) WAS REVIEWED FOR DEVIATIONS AND/ OR ANOMALIES WITH NO RELEVANT DEVIATIONS / ANOMALIES IDENTIFIED. ROOT CAUSE CANNOT BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL CONCOMITANT MEDICAL PRODUCTS: BMT 360 TIB TRAY 75MM, ITEM# 185204, LOT# 182190; BMT 360 TIB SM CRUCIATE WING, ITEM# 185650, LOT# 433600. CUSTOMER HAS INDICATED THAT THE PRODUCT WILL BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE OFFSET AND SMALL WING WAS ALREADY ENGAGED ON THE 75 PLATE AND COULD NOT BE DISENGAGED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.