FDA Adverse Event Malfunction Summary report: N

GMK-REVISION FIXED TIBIAL INSERT SC SIZE 3/20MM

MDR report key: 8050711 · Received November 8, 2018

Report

Report Number
3005180920-2018-00878
Event Type
Malfunction
Date Received
November 8, 2018
Date of Event
October 12, 2018
Report Date
December 19, 2018
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JWH
UDI-DI
07630030817519
PMA / PMN Number
K103170
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

VISUAL INSPECTION OF THE UHMWPE SEMI-CONSTRAINED REVISION INSERT: SOME DENTS AND SCRATCHES CAN BE NOTED ON THE DISTAL SURFACE OF THE INSERT. THE POSTERIOR FEATURES OF THE 'SNAPPING' MECHANISM HAVE BEEN PLASTICALLY DEFORMED AND DAMAGED, ESPECIALLY IN ONE SIDE. THEY BOTH PRESENT A SORT OF INCISION WITH THE NEGATIVE SHAPE OF THE CLIPPING "TOOTH" OF THE BASEPLATE. WE CAN SUPPOSE THAT, IN THE FIRST ATTEMPT TO FIX THE INSERT INTO THE BASEPLATE, THE INSERT WAS POSTERIORLY NOT WELL POSITIONED (WE CAN SUPPOSE THAT THE INSERTION WAS DIFFICULT BECAUSE THE JOINT WAS TIGHT USING A 20MM THICK INSERT; THAT'S THE REASON WHY WAS THEN USED A 17MM INSERT). IN THIS ATTEMPT THE INSERT WAS PUSHED POSTERIORILY AND WAS PERMANENTLY DAMAGED WITHOUT POSSIBILITY TO BE CLIPPED IN THE FOLLOWING ATTEMPTS. WE CAN STATE THAT THE EVENT IS NOT IMPLANT RELATED.

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 08 NOVEMBER 2018: LOT 140476: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 06-JUN-2014. EXPIRATION DATE: 2019-04-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

THE TIBIAL INSERT (22MM) WOULD NOT SNAP INTO THE TIBIAL TRAY. THE SURGEON ENDED UP DROPPING TO A 17MM INSERT AND THAT WORKED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
893140 GMK-REVISION FIXED TIBIAL INSERT SC SIZE 3/20MM KNEE TIBIAL INSERT JWH MEDACTA INTERNATIONAL SA 140476 07630030817519

Patients

Seq Age Sex Outcome Treatment
0 Other