FDA Adverse Event Injury Summary report: N

ATTUNE FB TIB BASE SZ 4 CEM

MDR report key: 7331452 · Received March 12, 2018

Report

Report Number
1818910-2018-54896
Event Type
Injury
Date Received
March 12, 2018
Date of Event
February 23, 2018
Report Date
February 23, 2018
Manufacturer
DEPUY IRELAND - 9616671
Product Code
JWH
UDI-DI
10603295042389
PMA / PMN Number
K101433
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT #: (B)(4). INVESTIGATION SUMMARY: NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE COMPLAINT DATABASE SEARCH FOUND NO OTHER REPORTED INCIDENT(S) AGAINST THE PROVIDED PRODUCT/LOT COMBINATION(S) SINCE RELEASE FOR DISTRIBUTION. CORRECTIVE ACTION WAS NOT INDICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

(B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

SURGEON: (B)(6), PRODUCT: ATTUNE, COMPLAINT OVERVIEW: LIFTING OF TIBIAL IMPLANT OUT OF THE PREPARED TIBIA DURING FINAL REDUCTION. PATIENT DETAILS: FEMALE, (B)(6), PRE OP OBSERVATIONS: HYPER EXTENSION IN KNEE. TWENTY (20) DEGREE FIXED VALGUS DEFORMITY. DETAILS OF PRODUCT COMPLAINT: EVENT TOOK PLACE WHILE CEMENTING IN FINAL IMPLANTS. KNEE PLACED IN FLEXION TO GAIN ACCESS TO IMPLANT COMPONENTS TIBIA SUBLUXED FORWARD TO ACCESS AND IMPLANT TIBIAL TRAY. ON REDUCTION, THE PREPARED FEMUR WOULD PUT FORCE ON THE POSTERIOR PART OF THE TIBIAL TRAY CAUSING THE TRAY TO LIFT OUT OF THE TIBIAL BONE. TIBIAL COMPONENT DROPPED ON THE FLOOR BY ACCIDENT. NEW TIBIAL COMPONENT OPENED. AN ADDITIONAL ASSISTANT WAS BROUGHT IN TO HELP WITH THE FINAL REDUCTION: ONE ASSISTANT TO PULL TIBIA FORWARD, ONE ASSISTANT TO LIFT FEMUR, SURGEON GUIDED REDUCTION WHILE KEEPING TIBIAL TRAY IN PLACE. FEMORAL COMPONENT IMPLANTED. TIBIAL INSERT TRIAL PLACED IN. KNEE PUT OUT TO EXTENSION TO ALLOW CEMENT TO SET. DEFINITIVE TIBIAL INSERT IMPLANTED. COMPONENT DETAILS: ATTUNE TIBIAL BASE SIZE 4, FIXED BEARING (1506-00-004). LOT: 8678183. (THIS COMPONENT WAS DROPPED AND NOT USED) ATTUNE TIBIAL BASE SIZE 4, FIXED BEARING (1506-00-004). LOT: 8611368. (COMPONENT IMPLANTED)

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
174917 ATTUNE FB TIB BASE SZ 4 CEM ATTUNE IMPLANT : KNEE TIBIAL TRAY JWH DEPUY IRELAND - 9616671 8678183 10603295042389

Patients

Seq Age Sex Outcome Treatment
1 66 YR Required Intervention