FDA Adverse Event Malfunction Summary report: N

PERSONA FIXED BEARING ARTICULAR SURFACE

MDR report key: 7132329 · Received December 19, 2017

Report

Report Number
3007963827-2017-00291
Event Type
Malfunction
Date Received
December 19, 2017
Date of Event
November 27, 2017
Report Date
March 29, 2018
Manufacturer
ZIMMER ORTHOPAEDIC MFG. LTD.
Product Code
JWH
PMA / PMN Number
PK113369
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

REPORTED EVENT WAS UNABLE TO BE CONFIRMED DUE TO LIMITED INFORMATION RECEIVED FROM THE CUSTOMER. DHR WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. REVIEW OF THE COMPLAINT HISTORY DETERMINED THAT NO FURTHER ACTION IS REQUIRED AS NO WERE TRENDS IDENTIFIED. THE ROOT CAUSE OF THE REPORTED ISSUE IS ATTRIBUTED TO USER ERROR. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT THIS TYPE OF EVENT CAN OCCUR AND RISKS ARE ADDRESSED IN THE ASSOCIATED RISK DOCUMENTATION. A SUMMARY OF THE INVESTIGATION WAS SENT TO THE COMPLAINANT CONVEYING PROPER SURGICAL TECHNIQUE AND USE OF THE DEVICE. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 1

ARTICULAR SURFACE WAS RETURNED FOR EVALUATION. VISUAL EVALUATION INDICATES THAT THE DOVE TAIL FEATURE AS WELL AS THE MEDIAL AND LATERAL RAILS ARE DAMAGED FLARED OUT. DIMENSIONAL EVALUATION INDICATES THAT THE DEVICE WAS MADE TO PRINT WHERE MEASURED. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT THIS TYPE OF EVENT CAN OCCUR AND RISKS ARE ADDRESSED IN RISK DOCUMENTATION. ROOT CAUSE CANNOT BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS

Additional Manufacturer Narrative · 1

(B)(4). REPORT SOURCE: (B)(4). CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION, AS PRODUCT LOCATION IS UNKNOWN. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ARTICULAR SURFACE WOULD NOT SEAT INTO THE TIBIAL BASE. THIS WAS REPEATED WITH THE SAME RESULTS. A NEW ARTICULAR SURFACE WAS USED TO COMPLETE THE CASE. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
911066 PERSONA FIXED BEARING ARTICULAR SURFACE PROSTHESIS, KNEE JWH ZIMMER ORTHOPAEDIC MFG. LTD. N/A 63793831

Patients

Seq Age Sex Outcome Treatment
1