PERSONA FIXED BEARING ARTICULAR SURFACE
Report
- Report Number
- 3007963827-2017-00291
- Event Type
- Malfunction
- Date Received
- December 19, 2017
- Date of Event
- November 27, 2017
- Report Date
- March 29, 2018
- Manufacturer
- ZIMMER ORTHOPAEDIC MFG. LTD.
- Product Code
- JWH
- PMA / PMN Number
- PK113369
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
REPORTED EVENT WAS UNABLE TO BE CONFIRMED DUE TO LIMITED INFORMATION RECEIVED FROM THE CUSTOMER. DHR WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. REVIEW OF THE COMPLAINT HISTORY DETERMINED THAT NO FURTHER ACTION IS REQUIRED AS NO WERE TRENDS IDENTIFIED. THE ROOT CAUSE OF THE REPORTED ISSUE IS ATTRIBUTED TO USER ERROR. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT THIS TYPE OF EVENT CAN OCCUR AND RISKS ARE ADDRESSED IN THE ASSOCIATED RISK DOCUMENTATION. A SUMMARY OF THE INVESTIGATION WAS SENT TO THE COMPLAINANT CONVEYING PROPER SURGICAL TECHNIQUE AND USE OF THE DEVICE. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
ARTICULAR SURFACE WAS RETURNED FOR EVALUATION. VISUAL EVALUATION INDICATES THAT THE DOVE TAIL FEATURE AS WELL AS THE MEDIAL AND LATERAL RAILS ARE DAMAGED FLARED OUT. DIMENSIONAL EVALUATION INDICATES THAT THE DEVICE WAS MADE TO PRINT WHERE MEASURED. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT THIS TYPE OF EVENT CAN OCCUR AND RISKS ARE ADDRESSED IN RISK DOCUMENTATION. ROOT CAUSE CANNOT BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS
(B)(4). REPORT SOURCE: (B)(4). CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION, AS PRODUCT LOCATION IS UNKNOWN. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT THE ARTICULAR SURFACE WOULD NOT SEAT INTO THE TIBIAL BASE. THIS WAS REPEATED WITH THE SAME RESULTS. A NEW ARTICULAR SURFACE WAS USED TO COMPLETE THE CASE. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 911066 | PERSONA FIXED BEARING ARTICULAR SURFACE | PROSTHESIS, KNEE | JWH | ZIMMER ORTHOPAEDIC MFG. LTD. | N/A | 63793831 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |