FDA Adverse Event Injury Summary report: N

GMK-SPHERE TIBIAL TRAY FIXED CEMENTED SIZE 2 RIGHT

MDR report key: 6944539 · Received October 12, 2017

Report

Report Number
3005180920-2017-00575
Event Type
Injury
Date Received
October 12, 2017
Date of Event
September 12, 2017
Report Date
March 23, 2018
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JWH
UDI-DI
07630030819872
PMA / PMN Number
K090988
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ASSOCIATED INSERT IS TIBIAL INSERT FIXED SPHERE FLEX # 2/17 MM R REFERENCE. THE POSSIBLE LOT. IS 140348 OR 120349. BATCH REVIEW PERFORMED ON 21 FEB 2018: LOT 140348, (B)(4) PIECES MANUFACTURER AND RELEASED ON 13 MARCH 2014. EXPIRATION DATE: 2019-01-31. NO ANOMALY FOUND RELATED TO THE ISSUE. TO DATE, (B)(4) PIECES OF THE SAME LOT HAS BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT. LOT 120349: (B)(4) PIECES MANUFACTURER AND RELEASED ON 02 MAY 2012. EXPIRATION DATE: 2017-03-31. NO ANOMALY FOUND RELATED TO THE ISSUE. TO DATE, (B)(4) PIECES OF THE SAME LOT HAS BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Additional Manufacturer Narrative · 1

THIS IS THE SECOND REVISION SURGERY UNDERWENT BY THE PATIENT. SHE HAD A PRIMARY SURGERY ON (B)(6) 2015. THE PATIENT FELL AND CAME IN COMPLAINING OF PAIN. THE SURGEON NOTICED THE PATELLA WAS LOOSE. ON (B)(6) 2017 THE SURGEON SWAPPED THE POLY AND PATELLA. THE SURGERY WAS COMPLETED SUCCESSFULLY. THEN THE PATIENT CAME IN COMPLAINING OF PAIN. THE SURGEON DETERMINED THAT THE TIBIA COMPONENT HAD SUBLUXED AND CAUSED A STRESS FRACTURE ON THE WALL OF HER TIBIA. BATCH REVIEW PERFORMED ON 12 OCTOBER 2017. (B)(4).

Description of Event or Problem · 1

THE PATIENT CAME IN COMPLAINING OF PAIN. THE SURGEON DETERMINED THAT THE TIBIA COMPONENT HAD SUBLUXED AND CAUSED A STRESS FRACTURE ON THE WALL OF HER TIBIA. THE SURGEON REVISED THE TIBIAL TRAY AND THE POLY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
724922 GMK-SPHERE TIBIAL TRAY FIXED CEMENTED SIZE 2 RIGHT TIBIAL TRAY FIXED CEMENTED JWH MEDACTA INTERNATIONAL SA 152800 07630030819872

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention