ROHS MICROSENSOR BASIC KIT
Report
- Report Number
- 1226348-2017-10525
- Event Type
- Malfunction
- Date Received
- July 31, 2017
- Date of Event
- July 7, 2017
- Manufacturer
- CODMAN & SHURTLEFF, INC.
- Product Code
- GWM
- PMA / PMN Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE 510(K) # OF SIMILAR PRODUCT CODE OF 826631: K914479. UPON COMPLETION OF THE INVESTIGATION A FOLLOW UP REPORT WILL BE FILED.
THE SENSOR WAS RETURNED AND EVALUATED BY THE SUPPLIER. A REVIEW OF QUALITY RECORDS WAS PERFORMED AND FOUND THAT THE COMPONENT MET ALL MANUFACTURING AND QUALITY TESTING/INSPECTION SPECIFICATIONS PRIOR TO DISTRIBUTION. EVALUATION OF THE RETURNED SENSOR FOUND THAT THE EPOXY BALL WAS PARTIALLY DISLODGED AT THE END OF THE SENSOR CASE. THIS IS NOT PART OF THE MANUFACTURING PROCESS. THE CATHETER MATERIAL WAS MELTED 8.5 CM FROM THE TIP OF THE SENSOR CASE AND KINKED 4.3 CM FROM THE TIP OF THE SENSOR CASE. THE DEVICE PASSED ELECTRONIC, NOISE, AND LINEARITY/HYSTERESIS TESTS. THE SENSOR FAILED THE DRIFT TEST AT DAY SEVEN OF THE TEST. IT IS SUSPECTED THAT THE EPOXY BALL SEPARATION CONTRIBUTED TO THESE RESULTS. THE DRIFT FAILURE, WHICH WAS OBSERVED ON A DAMAGED, USED SENSOR, WAS 1.6 MMHG OUT OF SPEC ON THE SEVENTH DAY OF TESTING AND WOULD HAVE MINIMAL EFFECT ON CUSTOMER'S USE. BASED ON THE SUPPLIER'S EVALUATION OF THE DEVICE, THE ISSUE REPORTED BY THE CUSTOMER COULD NOT BE CONFIRMED. TRENDS WILL BE MONITORED FOR THIS OR SIMILAR COMPLAINTS.
PROBE SHOWS AGAIN WRONG NUMBERS LIKE ALL THE OTHER COMPLAINT PROBES FROM THIS HOSPITAL SEND YOU BEFORE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 534637 | ROHS MICROSENSOR BASIC KIT | DEVICE, INTRACRANIAL PRESSURE MONITORING | GWM | CODMAN & SHURTLEFF, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |