FDA Adverse Event Injury Summary report: N

VANGUARD COMPLETE KNEE SYSTEM - E1 VNGD AS TIB BRG 10X83

MDR report key: 6578791 · Received May 19, 2017

Report

Report Number
0001825034-2017-03381
Event Type
Injury
Date Received
May 19, 2017
Date of Event
April 10, 2017
Report Date
September 12, 2017
Manufacturer
BIOMET ORTHOPEDICS
Product Code
JWH
PMA / PMN Number
PK113550
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT MEDICAL PRODUCTS ¿ REGENEREX PATELLA CATALOG 141357 LOT 344270; VANGUARD LEFT FEMUR 75MM CATALOG 183074 LOT 157830; BIOMET REGENEREX TIBIAL TRAY CATALOG 141276 LOT 572640; MODULAR SPLINED STEM 40MM CATALOG 141369 LOT 501970; CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORT: 1825034-2017-03380.

Additional Manufacturer Narrative · 1

COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS CONFIRMED. VISUAL INSPECTION OF THE RETURNED BEARING SHOWED WEAR MARKS ON THE PRODUCT. THE BEARING HAS FLAKES OF METAL FROM THE REGENEREX BACKING ON THE ARTICULAR SURFACE. DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. REVIEW OF THE COMPLAINT HISTORY DETERMINED THAT NO FURTHER ACTION IS REQUIRED. THE ROOT CAUSE OF THE BEARING WEAR IS ATTRIBUTED TO THIRD BODY WEAR AS A RESULT OF THE PATELLA PEG FRACTURING. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT A LEFT KNEE REVISION PROCEDURE APPROXIMATELY THREE YEARS POST IMPLANTATION DUE TO WEAR AND COMPONENT FRACTURE. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
360260 VANGUARD COMPLETE KNEE SYSTEM - E1 VNGD AS TIB BRG 10X83 PROSTHESIS, KNEE JWH BIOMET ORTHOPEDICS 802890

Patients

Seq Age Sex Outcome Treatment
1 76 YR Hospitalization| R