VANGUARD COMPLETE KNEE SYSTEM - E1 VNGD AS TIB BRG 10X83
Report
- Report Number
- 0001825034-2017-03381
- Event Type
- Injury
- Date Received
- May 19, 2017
- Date of Event
- April 10, 2017
- Report Date
- September 12, 2017
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- JWH
- PMA / PMN Number
- PK113550
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). CONCOMITANT MEDICAL PRODUCTS ¿ REGENEREX PATELLA CATALOG 141357 LOT 344270; VANGUARD LEFT FEMUR 75MM CATALOG 183074 LOT 157830; BIOMET REGENEREX TIBIAL TRAY CATALOG 141276 LOT 572640; MODULAR SPLINED STEM 40MM CATALOG 141369 LOT 501970; CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORT: 1825034-2017-03380.
COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS CONFIRMED. VISUAL INSPECTION OF THE RETURNED BEARING SHOWED WEAR MARKS ON THE PRODUCT. THE BEARING HAS FLAKES OF METAL FROM THE REGENEREX BACKING ON THE ARTICULAR SURFACE. DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. REVIEW OF THE COMPLAINT HISTORY DETERMINED THAT NO FURTHER ACTION IS REQUIRED. THE ROOT CAUSE OF THE BEARING WEAR IS ATTRIBUTED TO THIRD BODY WEAR AS A RESULT OF THE PATELLA PEG FRACTURING. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT THE PATIENT UNDERWENT A LEFT KNEE REVISION PROCEDURE APPROXIMATELY THREE YEARS POST IMPLANTATION DUE TO WEAR AND COMPONENT FRACTURE. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 360260 | VANGUARD COMPLETE KNEE SYSTEM - E1 VNGD AS TIB BRG 10X83 | PROSTHESIS, KNEE | JWH | BIOMET ORTHOPEDICS | 802890 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Hospitalization| R |