FDA Adverse Event Malfunction Summary report: N

SUBDURAL POST CRANIOTOMY ICP MONITORING KIT

MDR report key: 6311737 · Received February 8, 2017

Report

Report Number
2023988-2017-00012
Event Type
Malfunction
Date Received
February 8, 2017
Report Date
January 19, 2017
Manufacturer
INTEGRA NEUROSCIENCES CA/USA
Product Code
GWM
PMA / PMN Number
K102875
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IT IS UNKNOWN IF THE DEVICE INVOLVED IN THE REPORTED INCIDENT IS EXPECTED TO BE RETURNED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED UPON THE REPORTED INFORMATION.

Additional Manufacturer Narrative · 1

INTEGRA HAS COMPLETED THEIR INTERNAL INVESTIGATION ON MARCH 16, 2017. THE INVESTIGATION INCLUDED: METHODS: REVIEW OF DEVICE HISTORY RECORDS. REVIEW OF COMPLAINTS HISTORY. RESULTS: THE CATHETER HAS NOT BEEN RETURNED FOR INVESTIGATION. DHR REVIEW; THERE WERE (B)(4) LOTS MODEL 110-4L SHIPPED TO THE CUSTOMER FROM JANUARY 1ST 2016 TO JANUARY 15TH 2017; THEIR BATCH HISTORY RECORDS INDICATING THAT ALL REQUIREMENTS MET BEFORE THEY WERE RELEASED TO FINISHED GOODS. COMPLAINTS HISTORY; (B)(4). CONCLUSION: COULD NOT BE CONFIRM SINCE THE CUSTOMER DID NOT RETURN THE DEVICE FOR EVALUATION.

Description of Event or Problem · 1

THE PROBES ARE GIVING ERRATIC READINGS, FROM -5 TO 145 AND BACK TO 5 WHILE THE SURGEON IS TALKING TO THE PATIENT AND WATCHING THE MONITOR. THEY HAVE HAD FOUR 1104G PROBES WITH ISSUES IN THIS MONTH. THE HOSPITAL S A REGULAR USER AND IS VERY AWARE OF HOW THE PRODUCT MUST BE USED. DISTRIBUTOR'S TECHNICIAN CHECKED THE MONITORS AND THEY ARE WORKING PERFECTLY. ADDITIONAL INFORMATION HAS BEEN REQUESTED. LINKED TO MFG. REPORT NUMBERS: 2023988-2017-00011, 2023988-2017-00013, 2023988-2017-00014, 2023988-2017-00015, 2023988-2017-00016, 2023988-2017-00017, 2023988-2017-00018.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
93388 SUBDURAL POST CRANIOTOMY ICP MONITORING KIT N/A GWM INTEGRA NEUROSCIENCES CA/USA

Patients

Seq Age Sex Outcome Treatment
1 CAMINO MONITOR