TRIMA ACCEL
Report
- Report Number
- 1722028-2016-00616
- Event Type
- Malfunction
- Date Received
- November 29, 2016
- Date of Event
- November 2, 2016
- Report Date
- November 29, 2016
- Manufacturer
- TERUMO BCT
- Product Code
- GKT
- PMA / PMN Number
- BK150321
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
INVESTIGATION: THE DEVICE HISTORY RECORDS (DHR) WERE REVIEWED FOR THIS LOT. THERE WERE NO EVENTS NOTED IN THE DHR THAT WOULD HAVE CONTRIBUTED TO THE ELEVATED WBC COUNT EXPERIENCED BY THE CUSTOMER. THE RUN DATA FILE (RDF) WAS ANALYZED FOR THIS EVENT. ROOT CAUSE: A DEFINITIVE ROOT CAUSE FOR THE OBSERVED LEUKOREDUCTION FAILURE REMAINS UNDETERMINED AT THIS TIME. THE ANALYSIS OF THE RUN DATA FILE DID NOT FIND A CONCLUSIVE CAUSE FOR THE HIGHER-THAN-EXPECTED WBC CONTENT IN THE PLATELET PRODUCT REPORTED FOR THIS COLLECTION. NO UNUSUAL PROCESS VARIABLE WAS IDENTIFIED AND THE SIGNALS IN THE RUN DATA FILE INDICATE THAT THE TRIMA ACCEL SYSTEM OPERATED AS INTENDED. BASED ON THE AVAILABLE INFORMATION, IT CANNOT BE RULED OUT THAT THE HIGHER-THAN-EXPECTED WBC CONTENT IN THE PLATELET PRODUCT COULD BE DONOR-RELATED. IT ALSO CANNOT BE RULED OUT THAT A SAMPLING, CALCULATION, OR OTHER PROCESS ERROR COULD HAVE CONTRIBUTED TO THE HIGHER-THAN-EXPECTED WBC CONTENT IN THE PLATELET PRODUCT.
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE PLATELET PRODUCT. THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WHITE BLOOD CELL (RWBC) TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. DONOR UNIT #:(B)(6). THE PLATELET COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 783352 | TRIMA ACCEL | TRIMA ACCEL PLATELET + SAMPLER, PLASMA RBC SET | GKT | TERUMO BCT | 09Z4213 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00000 | Other |