FDA Adverse Event Malfunction Summary report: N

TRIMA ACCEL

MDR report key: 6110567 · Received November 17, 2016

Report

Report Number
1722028-2016-00606
Event Type
Malfunction
Date Received
November 17, 2016
Date of Event
September 15, 2016
Report Date
November 17, 2016
Manufacturer
TERUMO BCT
Product Code
GKT
PMA / PMN Number
BK150321
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: THE DEVICE HISTORY RECORDS (DHR) WERE REVIEWED FOR THIS LOT. THERE WERE NO EVENTS NOTED IN THE DHR THAT WOULD HAVE CONTRIBUTED TO THE ELEVATED WBC COUNT EXPERIENCED BY THE CUSTOMER. ROOT CAUSE: A DEFINITIVE ROOT CAUSE FOR THE OBSERVED LEUKO REDUCTION FAILURE REMAINS UNDETERMINED AT THIS TIME. THE ANALYSIS OF THE RUN DATA FILE DID NOT FIND A CONCLUSIVE CAUSE FOR THE HIGHER-THAN-EXPECTED WBC CONTENT IN THE PLATELET PRODUCT REPORTED FOR THIS COLLECTION. NO UNUSUAL PROCESS VARIABLE WAS IDENTIFIED AND THE SIGNALS IN THE RUN DATA FILE INDICATE THAT THE TRIMA ACCEL SYSTEM OPERATED AS INTENDED.

Additional Manufacturer Narrative · 1

INVESTIGATION: THE RUN DATA FILE (RDF) WAS ANALYZED FOR THIS EVENT. THE ANALYSIS OF THE RUN DATA FILE DID NOT FIND A CONCLUSIVE CAUSE FOR THE HIGHER-THAN-EXPECTED WBC CONTENT IN THE PLATELET PRODUCT REPORTED FOR THIS COLLECTION. NO UNUSUAL PROCESS VARIABLE WAS IDENTIFIED AND THE SIGNALS IN THE RUN DATA FILE INDICATE THAT THE TRIMA ACCEL SYSTEM OPERATED AS INTENDED. INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED.

Description of Event or Problem · 1

THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE PLATELET PRODUCT. THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WHITE BLOOD CELL (RWBC) TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. DONOR UNIT #: (B)(4). PLATELET COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
762297 TRIMA ACCEL TRIMA ACCEL ENHANCED PLATELET, PLASMA, RBC SET GKT TERUMO BCT 08Z4115

Patients

Seq Age Sex Outcome Treatment
1 Other