FDA Adverse Event Injury Summary report: N

NEXGEN MIS DROP DOWN STEM PLUG

MDR report key: 6070533 · Received November 1, 2016

Report

Report Number
0001822565-2016-03751
Event Type
Injury
Date Received
November 1, 2016
Date of Event
December 19, 2014
Report Date
November 29, 2016
Manufacturer
ZIMMER, INC.
Product Code
JWH
PMA / PMN Number
PK043101
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION THAT WAS NOT AVAILABLE AT THE TIME THAT THE INITIAL MEDWATCH WAS SUBMITTED. CONCOMITANT PRODUCTS: NEXGEN LPS-FLEX GSF FEMORAL COMPONENT OPTION, CATALOGUE # 00576401552, LOT # 60779312. NEXGEN MIS STEMMED TIBIAL COMPONENT, CATALOGUE # 00595003702, LOT #60616289. NEXGEN LPS-FLEX ARTICULAR SURFACE, CATALOGUE # 00-5962-032-14 LOT #:60512473. NEXGEN ALL POLY PATELLA, CATALOGUE # 00597206529, LOT # 60759304. PALACOS BONE CEMENT, CATALOGUE # 00111314001, LOT # 65584099. NO PRODUCT WAS RETURNED; VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. THE DEVICE HISTORY RECORDS FOR THE STEM EXTENSION WERE REVIEWED AND NO DEVIATIONS OR ANOMALIES WERE IDENTIFIED. REVIEW OF THE COMPATIBILITY MATRIX IDENTIFIED THAT ALL COMPONENTS USED WERE COMPATIBLE. THE DEVICE IS USED FOR TREATMENT. THE PRIMARY SURGICAL NOTES STATE THAT THE PATIENT UNDERWENT RIGHT TKA DUE TO SEVERE ARTHRITIS IN THE RIGHT KNEE. THE BONE CUTS WERE COMPLETED, THE MENISCI AND PCL WERE COMPLETELY EXCISED. THE TIBIAL CUT WAS MADE PERPENDICULAR TO THE 2ND RAY OF THE FOOT AND CHECKED WITH MULTIPLE ALIGNMENT RODS TO INSURE ACCURACY. A TRIAL REDUCTION WAS PERFORMED AND EXCELLENT STABILITY THROUGH ALL RANGES WAS INITIALLY ACHIEVED WITH A 14MM POLY. THE FINAL COMPONENTS WERE CEMENTED IN PLACE IN A NORMAL FASHION. EXCESS CEMENT WAS REMOVED. PULSATILE LAVAGE WITH THOROUGH IRRIGATION WAS USED ONCE AGAIN BEFORE WOUND CLOSURE. THE RANGE OF MOTION AND STABILITY WERE BOTH EXCELLENT. THE REVISION SURGICAL NOTES INDICATE THAT RADIOGRAPHS SHOWED OBVIOUS LOOSENING OF THE TIBIAL COMPONENT AND ABOUT 30 DEGREES OF VARUS. THERE WAS NO EVIDENCE OF INFECTION ON FROZEN SECTION AND NO FRANK PURULENCE. INTRA-OPERATIVELY, THE TIBIAL COMPONENT WAS FOUND TO BE OBVIOUSLY LOOSE AND REMOVED WITH A POWER OSTEOTOME AND BONE TAMP. THERE APPEARED TO BE CYSTIC DEGENERATION OF THE MEDIAL ASPECT OF THE TIBIAL PLATEAU IN THE VICINITY OF SUBSIDENCE, WHICH RESULTED IN VARUS DEFORMITY. THE FEMORAL COMPONENT WAS ALSO LOOSENED IN A SIMILAR FASHION. PER THE PACKAGE INSERTS OF THE COMPONENTS, PAIN AND LOOSENING OF THE PROSTHETIC KNEE COMPONENTS ARE KNOWN POTENTIAL ADVERSE EFFECTS OF THE TKA PROCEDURE. THE INSERT ALSO WARNS THAT COMPLICATIONS AND/OR FAILURE OF TOTAL KNEE PROSTHESES ARE MORE LIKELY TO OCCUR IN HEAVY PATIENTS. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED WITH THE INFORMATION PROVIDED.

Additional Manufacturer Narrative · 1

THIS REPORT WILL BE AMENDED WHEN OUR INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

IT HAS BEEN REPORTED THAT FOLLOWING A KNEE ARTHROPLASTY, THE PATIENT WAS REVISED DUE TO PAIN AND TIBIAL LOOSENING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
720969 NEXGEN MIS DROP DOWN STEM PLUG KNEE PROSTHESIS JWH ZIMMER, INC. 60411831

Patients

Seq Age Sex Outcome Treatment
1 55 YR Hospitalization| R