FDA Adverse Event Malfunction Summary report: N

TRIMA ACCEL

MDR report key: 6037681 · Received October 18, 2016

Report

Report Number
1722028-2016-00571
Event Type
Malfunction
Date Received
October 18, 2016
Date of Event
September 29, 2016
Report Date
October 18, 2016
Manufacturer
TERUMO BCT
Product Code
GKT
PMA / PMN Number
BK150321
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: THE DEVICE HISTORY RECORDS (DHR) WERE REVIEWED FOR THIS LOT. THERE WERE NO EVENTS NOTED IN THE DHR THAT WOULD HAVE CONTRIBUTED TO THE ELEVATED WBC COUNT EXPERIENCED BY THE CUSTOMER. THE RUN DATA FILE (RDF) WAS ANALYZED FOR THIS EVENT. ROOT CAUSE: A DEFINITIVE ROOT CAUSE FOR THE OBSERVED LEUKOREDUCTION FAILURE REMAINS UNDETERMINED AT THIS TIME. THE ANALYSIS OF THE RUN DATA FILE DID NOT FIND A CONCLUSIVE CAUSE FOR THE HIGHER-THAN-EXPECTED WBC CONTENT IN THE PLATELET PRODUCT REPORTED FOR THIS COLLECTION. NO UNUSUAL PROCESS VARIABLE WAS IDENTIFIED AND THE SIGNALS IN THE RUN DATA FILE INDICATE THAT THE TRIMA ACCEL SYSTEM OPERATED AS INTENDED. BASED ON THE AVAILABLE INFORMATION, IT CANNOT BE RULED OUT THAT THE HIGHER-THAN-EXPECTED WBC CONTENT IN THE PLATELET PRODUCT COULD BE DONOR-RELATED. IT ALSO CANNOT BE RULED OUT THAT A SAMPLING, CALCULATION, OR OTHER PROCESS ERROR COULD HAVE CONTRIBUTED TO THE HIGHER-THAN-EXPECTED WBC CONTENT IN THE PLATELET PRODUCT.

Description of Event or Problem · 1

THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE PLATELET PRODUCT. THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WHITE BLOOD CELL (RWBC) TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. DONOR UNIT #: (B)(6) THE PLATELET COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
687268 TRIMA ACCEL TRIMA ACCEL PLATELET, PLASMA, RBC SET GKT TERUMO BCT 08Z3220

Patients

Seq Age Sex Outcome Treatment
1 Other