FDA Adverse Event Malfunction Summary report: N

ECARECOORDINATOR

MDR report key: 6028657 · Received October 13, 2016

Report

Report Number
1125873-2016-00042
Event Type
Malfunction
Date Received
October 13, 2016
Report Date
October 5, 2016
Manufacturer
VISICU, INC
Product Code
DRG
PMA / PMN Number
K141706
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

PHILIPS Q&R HAS COMPLETED THE INVESTIGATION FOR THE REPORTED FAILURE OF PROTOCOLS ENDING WITHOUT NOTICE (RNS ARE UNAWARE) WITH THE ECARE COMPANION HOME TELEMONITORING SYSTEM. THIS COMPLAINT HAS BEEN CONFIRMED AND A FIX WAS IMPLEMENTED VIA FCO 1125873-10-06-16-017-C IN SOFTWARE VERSION 1.4.0.5. Q&R WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 1

IF A PATIENT GENERATES AN ADHERENCE FLAG ON THE FINAL DAY OF THE PATIENT'S IR, THE PATIENT'S INTERVENTION EXPIRED FIELD IS NOT BEING SET, AND NO CHECK PATIENT CALENDAR TASK WILL BE GENERATED FOR THE PATIENT. THIS ISSUE OCCURS IF YOU HAVE A PATIENT WHOSE INTERVENTION RULES EXPIRE AND, WHILE WAITING FOR HE TIME BASED ACTION TO EXECUTE, IF THE PATIENT RECORD IS UPDATED THEN THE PATIENT RECORD IS REMOVED FROM THE TIME-BASE WORKFLOW QUEUE. THIS CAUSES THE LAST INTERVENTION EXPIRED FIELD NOT TO GET UPDATES ON THE PATIENT AND SINCE THAT FIELD IS NOT SET ON THE PATIENT RECORD THEN THE CHECK PATIENT CALENDAR TASK IS NOT GENERATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
677450 ECARECOORDINATOR ECARECOORDINATOR DRG VISICU, INC 453564506091

Patients

Seq Age Sex Outcome Treatment
1