Tool · Post-Market Surveillance

PSUR Report

The vigilance database section of your PSUR, generated in seconds: FDA MAUDE adverse event trends, recalls, and enforcement classifications for your device and similar devices — with the documented, reproducible methodology auditors ask for.

Period
from
to
Leave empty for the last 24 months

Reporting period 2024-072026-06 · FDA data refreshed ~monthly

Product Code: DRG FDA class 2

Transmitters And Receivers, Physiological Signal, Radiofrequency

View full classification →
Adverse events in period
96
+96% vs. prior period (49)
Deaths reported
0
Recalls in period
1
Class I enforcement
0

Adverse events per month

2024-07 – 2026-06
Loading...

Events by type

Period vs. prior period
Loading...
Event type
Period
Prior
Injury
3
0
Malfunction
93
47
Death
0
2

Most reported coded problems

Top 15
Product problems
Count
Communication or Transmission Problem
67
Intermittent Communication Failure
12
Unable to Obtain Readings
5
Output Problem
5
Wireless Communication Problem
4
Incorrect, Inadequate or Imprecise Result or Readings
3
Computer System Security Problem
3
Power Problem
2
No Display/Image
2
Melted
2
Excessive Heating
2
Display or Visual Feedback Problem
2
Computer Software Problem
2
Smoking
1
Signal Artifact/Noise
1
Patient problems
Count
No Clinical Signs, Symptoms or Conditions
91
Insufficient Information
2
Urticaria
1
Sleep Dysfunction
1
Nodule
1
Itching Sensation
1
Easy Bruising
1
Burn(s)
1

Recalls in period

1 total
FDA enforcement classification: Class II: 1
Date
Recalling firm
Status
2024-07-19
Open, Classified

Adverse events by year

Loading...

Methodology & data provenance

For your PSUR appendix

Search protocol: FDA MAUDE adverse event reports, FDA device recalls, and FDA enforcement reports (openFDA datasets) were queried for product code DRG, reporting period 2024-07 to 2026-06 (prior comparison period 2022-07 to 2024-06).

Data coverage: adverse event data for these product codes extends through 2026-05. Source openFDA export dated 2026-06-30. Recall records with missing or implausible initiation dates (0 for these codes) are excluded from period counts.

Generated: 2026-07-06 16:46 UTC by BEUDAMED (beudamed.com/fda/psur-report). Dataset sizes: 25,039,198 adverse events, 58,607 recalls, 39,365 enforcement reports.

PSUR database searches, explained

Under the EU MDR, manufacturers must produce a Periodic Safety Update Report (PSUR) (Class IIa and above; a PMS report for Class I) on a defined schedule. A required part of that report is a survey of publicly available vigilance data — most importantly the FDA's MAUDE adverse event database and FDA recall records — covering both your own device and similar devices on the market (the "state of the art").

Doing this by hand means repeated MAUDE queries, manual export, deduplication, and trending — hours of work per report, repeated every reporting period. This tool runs the same search protocol in seconds and documents it: which datasets were queried, with which filters, when, and with what coverage — so the search is reproducible, which is exactly what a Notified Body auditor wants to see.

Event counts are grouped by the FDA's reported event type (death, injury, malfunction). Name-based device matching is clearly labeled as indicative: only you can confirm whether a specific MAUDE report concerns your device. Recall records with implausible dates are excluded and the exclusion is disclosed. All data is from the official openFDA datasets and refreshed approximately monthly.