LICOX OXYGEN MICRO-PROBE G-WM
Report
- Report Number
- 9617494-2004-00025
- Event Type
- Malfunction
- Date Received
- September 17, 2004
- Report Date
- September 17, 2004
- Manufacturer
- GESELLSCHAFT FUR MEDIZINISCHE SONDENTECHNIK GMS, M
- Product Code
- GWM
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
TO DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION.
PHYSICIAN PLACED THE IM3 BOLT. THE TEMPERATURE PROBE WAS PLACED WITHOUT INCIDENT, HOWEVER, THE OXYGEN PROBE WAS VERY DIFFICULT TO PLACE DOWN THE CHANNEL AND THE PHYSICIAN STRUGGLED TO GET THE PROBE IN PLACE. THE INITIAL OXYGEN READING WAS 0.5MM AND STAYED CONSISTENT FOR 3 HOURS. AN OXYGEN CHALLENGE WAS PERFORMED BUT THE NUMBER ONLY WENT UP TO 1MMHG. IT WAS OBVIOUS THAT THE PROBE WAS NOT WORKING CORRECTLY. ADDITIONAL MANIPULATION OF THE PROBE TO PLACE IT DOWN THE CHANNEL WAS REQUIRED. THE SYSTEM WAS REMOVED AFTER ABOUT FIVE HOURS IN PLACE WITHOUT ACCURATE OXYGEN READINGS. A NEW SYSTEM WAS PLACED IN THE SAME BURR HOLE THAT HAD BEEN MADE IN THE PATIENT'S HEAD. THE NEW OXYGEN PROBE IS READING CORRECTLY AT APPROXIMATELY 30MMHG.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LICOX OXYGEN MICRO-PROBE G-WM | LICOX BRAIN TISSUE OXYGEN MONITORING SYS | GWM | GESELLSCHAFT FUR MEDIZINISCHE SONDENTECHNIK GMS, M | 100504 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |