GMK PRIMARY FEMUR CEMENTED PS #5NARROW / LEFT
Report
- Report Number
- 3005180920-2015-00287
- Event Type
- Injury
- Date Received
- November 25, 2015
- Date of Event
- October 22, 2015
- Report Date
- February 18, 2016
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- JWH
- PMA / PMN Number
- K122232
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
ON 04 NOV 2015 IT WAS COMMUNICATED THAT THE RESULTS FOR THE PATHOGEN CAME BACK NEGATIVE. BATCH REVIEW PERFORMED ON 23 NOVEMBER 2015: GMK FEMUR CEMENTED PS #5NARROW / LEFT, CODE 02.07.2215L LOT. 150301: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 13 MARCH 2015. EXPIRATION DATE: 2020-01-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. GMK TIBIAL TRAY FIXED CEMENTED # 4 R, CODE 02.07.1204L LOT. 150912 (K090988): (B)(4) ITEMS MANUFACTURED AND RELEASED ON 12 MAY 2015. EXPIRATION DATE: 2020-04-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. GMK TIBIAL INSERT PS FIXED # 4/12MM, CODE 02.07.0412PSF LOT. 136094 (K090988): (B)(4) ITEMS MANUFACTURED AND RELEASED ON 10 FEBRUARY 2014. EXPIRATION DATE: 2018-12-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. GMK PATELLA RESURFACING # 2, CODE 02.07.0034RP LOT. 151408 (K090988): LOT 151408: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 21 AUGUST 2015. EXPIRATION DATE: 2020-06-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.
ON 22 JANUARY 2016, IT WAS PREPARED A FINAL REPORT WITH THE INFORMATION ALREADY SUBMITTED IN THE INTIAL REPORT. ON 25 JANUARY 2016, THE REPORT WAS SENT TO THE INITIAL REPORTER AND THE CASE WAS CLOSED.
ON (B)(6) 2015 PATIENT CAME IN FOR 6 WEEK POST-OPERATIVE CHECK-UP COMPLAINING OF PAIN AND INFLAMMATION. SURGEON TOOK A BLOOD SAMPLE WHICH SHOWED A HIGH LEVEL OF WHITE BLOOD CELLS. ON (B)(6) 2015 THE SURGEON REMOVED ALL COMPONENTS AND PUT IN AN ALL POLY COMPONENTS WITH ANTIBIOTIC CEMENT. THE EXPLANTS WILL NOT BE RETURNED. X-RAYS ARE NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 780616 | GMK PRIMARY FEMUR CEMENTED PS #5NARROW / LEFT | FEMORAL CEMENTED COMPONENT | JWH | MEDACTA INTERNATIONAL SA | 150301 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Required Intervention |