FDA Adverse Event Injury Summary report: N

GMK RESURFACING PATELLA SIZE 2

MDR report key: 5239222 · Received November 20, 2015

Report

Report Number
3005180920-2015-00277
Event Type
Injury
Date Received
November 20, 2015
Date of Event
October 21, 2015
Report Date
February 17, 2016
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JWH
PMA / PMN Number
K090988
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON 20 JANUARY 2016, IT WAS PREPARED A FINAL REPORT WITH THE INFORMATION ALREADY SUBMITTED IN THE INITIAL REPORT. ON 21 JANUARY 2016, THE REPORT WAS SENT TO THE INITIAL REPORTER AND THE CASE WAS CLOSED.

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON (B)(6) 2015: LOT 150873: (B)(4) ITEMS MANUFACTURED AND RELEASED ON (B)(6) 2015. EXPIRATION DATE: 2020-05-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. NOT AVAILABLE.

Description of Event or Problem · 1

TWO WEEKS POST-OPERATIVE THE PATIENT WAS UNDERGOING REHABILITATION. WHILE USING THE CONTINUOUS PASSIVE MOTION MACHINE THE PATIENT FELT A POP ON THE MEDIAL SIDE OF HIS KNEE. THE PATIENT DID NOT REPORT HIS EXPERIENCE TO THE SURGEON. THE SUTURE BROKE AND THE PATELLA LATERALIZED AND DISLOCATED. WHEN THE PATIENT NOTIFIED THE SURGEON OF THE PROBLEM HE DID A LATERAL RELEASE, VASTUS MEDIALIS OBLIQUE AND A MEDIAL CAPSULAR PLICATION. THE #2 PATELLA WAS REPLACED WITH A #1 PATELLA. THE SURGERY WAS COMPLETED SUCCESSFULLY. X-RAYS ARE NOT AVAILABLE. THE EXPLANT WILL NOT BE RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
770272 GMK RESURFACING PATELLA SIZE 2 RESURFACING PATELLA JWH MEDACTA INTERNATIONAL SA 150873

Patients

Seq Age Sex Outcome Treatment
1 58 YR Required Intervention