GMK SPHERE FLEX TIBIAL INSERT SIZE 4 / 10MM LEFT
Report
- Report Number
- 3005180920-2015-00245
- Event Type
- Malfunction
- Date Received
- October 24, 2015
- Date of Event
- September 24, 2015
- Report Date
- October 23, 2015
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- JWH
- PMA / PMN Number
- K121416
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
BATCH REVIEW PERFORMED ON 16 OCTOBER 2015: LOT 147252: 75 ITEMS MANUFACTURED AND RELEASED ON 26 MARCH 2015. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED ISSUE. GMK TIBIAL TRAY FIXED CEMENTED CODE 02.07.1204L # 4 R LOT. 150912 (K090988): BATCH REVIEW PERFORMED ON 20 OCTOBER 2015: LOT 150912: 35 ITEMS MANUFACTURED AND RELEASED ON 12 MAY 2015. EXPIRATION DATE: 2020-04-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 10 ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED ISSUE. INSPECTION OF THE RETRIEVED ITEM DONE ON THE 15TH OF OCTOBER 2015 BY (B)(4) PROJECT MANAGER: DURING THE VISUAL INSPECTION OF THE INSERT WAS NOTED THAT THE POSTERIOR "CLIPPING" FEATURES OF THE INSERT HAVE BEEN PLASTICALLY DEFORMED AND DAMAGED BOTH IN THE MEDIAL AND THE LATERAL SIDE. ONE OF THIS FEATURE PRESENT A SORT OF INCISION WITH THE NEGATIVE SHAPE OF THE CLIPPING "TOOTH" OF THE BASEPLATE. WE CAN SUPPOSE THAT PROBABLY, IN THE FIRST ATTEMPT TO FIX THE INSERT INTO THE BASEPLATE, THE INSERT WAS NOT POSTERIORLY WELL POSITIONED. IN THIS ATTEMPT THE INSERT WAS PERMANENTLY DAMAGED WITHOUT POSSIBILITY TO BE CLIPPED IN THE FOLLOWING ONES. WE CAN STATE THAT THE EVENT IS NOT IMPLANT RELATED. ON 21 OCTOBER 2015 IT WAS PREPARED A FINAL REPORT WITH THE INFORMATION ABOVE DESCRIBED. ON THE SAME DAY IT WAS SENT TO THE INITIAL REPORTER AND THE CASE WAS CLOSED.
DURING SURGERY THE SURGEON COULD NOT GET THE TIBIAL INSERT TO SEAT ON THE TIBIAL BASEPLATE. AFTER 15 MINUTES OF ADJUSTING THE SURGEON REQUESTED A NEW TIBIAL INSERT. THE NEW LINER SAT IMMEDIATELY AND THE SURGERY WAS COMPLETED SUCCESSFULLY. EXPLANT WILL BE RETURNED NO X-RAYS ARE AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 705854 | GMK SPHERE FLEX TIBIAL INSERT SIZE 4 / 10MM LEFT | FIXED PE TIBIAL INSERT | JWH | MEDACTA INTERNATIONAL SA | 147252 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 YR |